Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies.

Details

Ressource 1Download: 60.pdf (259.50 [Ko])
State: Public
Version: Final published version
License: Not specified
It was possible to publish this article open access thanks to a Swiss National Licence with the publisher.
Serval ID
serval:BIB_C07A7311AAB9
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies.
Journal
Neuroradiology
Author(s)
Komlosi P., Zhang Y., Leiva-Salinas C., Ornan D., Patrie J.T., Xin W., Grady D., Wintermark M.
ISSN
1432-1920 (Electronic)
ISSN-L
0028-3940
Publication state
Published
Issued date
2014
Peer-reviewed
Oui
Volume
56
Number
3
Pages
187-193
Language
english
Notes
Publication types: Journal Article ; Research Support, Non-U.S. Gov't Publication Status: ppublish
Abstract
INTRODUCTION: Adaptive statistical iterative reconstruction (ASIR) can decrease image noise, thereby generating CT images of comparable diagnostic quality with less radiation. The purpose of this study is to quantify the effect of systematic use of ASIR versus filtered back projection (FBP) for neuroradiology CT protocols on patients' radiation dose and image quality.
METHODS: We evaluated the effect of ASIR on six types of neuroradiologic CT studies: adult and pediatric unenhanced head CT, adult cervical spine CT, adult cervical and intracranial CT angiography, adult soft tissue neck CT with contrast, and adult lumbar spine CT. For each type of CT study, two groups of 100 consecutive studies were retrospectively reviewed: 100 studies performed with FBP and 100 studies performed with ASIR/FBP blending factor of 40 %/60 % with appropriate noise indices. The weighted volume CT dose index (CTDIvol), dose-length product (DLP) and noise were recorded. Each study was also reviewed for image quality by two reviewers. Continuous and categorical variables were compared by t test and free permutation test, respectively.
RESULTS: For adult unenhanced brain CT, CT cervical myelography, cervical and intracranial CT angiography and lumbar spine CT both CTDIvol and DLP were lowered by up to 10.9 % (p < 0.001), 17.9 % (p = 0.005), 20.9 % (p < 0.001), and 21.7 % (p = 0.001), respectively, by using ASIR compared with FBP alone. Image quality and noise were similar for both FBP and ASIR.
CONCLUSION: We recommend routine use of iterative reconstruction for neuroradiology CT examinations because this approach affords a significant dose reduction while preserving image quality.
Pubmed
Web of science
Create date
11/04/2014 18:42
Last modification date
09/09/2021 7:13
Usage data