Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies.

Détails

Ressource 1Télécharger: 60.pdf (259.50 [Ko])
Etat: Public
Version: Final published version
Licence: Non spécifiée
It was possible to publish this article open access thanks to a Swiss National Licence with the publisher.
ID Serval
serval:BIB_C07A7311AAB9
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies.
Périodique
Neuroradiology
Auteur⸱e⸱s
Komlosi P., Zhang Y., Leiva-Salinas C., Ornan D., Patrie J.T., Xin W., Grady D., Wintermark M.
ISSN
1432-1920 (Electronic)
ISSN-L
0028-3940
Statut éditorial
Publié
Date de publication
2014
Peer-reviewed
Oui
Volume
56
Numéro
3
Pages
187-193
Langue
anglais
Notes
Publication types: Journal Article ; Research Support, Non-U.S. Gov't Publication Status: ppublish
Résumé
INTRODUCTION: Adaptive statistical iterative reconstruction (ASIR) can decrease image noise, thereby generating CT images of comparable diagnostic quality with less radiation. The purpose of this study is to quantify the effect of systematic use of ASIR versus filtered back projection (FBP) for neuroradiology CT protocols on patients' radiation dose and image quality.
METHODS: We evaluated the effect of ASIR on six types of neuroradiologic CT studies: adult and pediatric unenhanced head CT, adult cervical spine CT, adult cervical and intracranial CT angiography, adult soft tissue neck CT with contrast, and adult lumbar spine CT. For each type of CT study, two groups of 100 consecutive studies were retrospectively reviewed: 100 studies performed with FBP and 100 studies performed with ASIR/FBP blending factor of 40 %/60 % with appropriate noise indices. The weighted volume CT dose index (CTDIvol), dose-length product (DLP) and noise were recorded. Each study was also reviewed for image quality by two reviewers. Continuous and categorical variables were compared by t test and free permutation test, respectively.
RESULTS: For adult unenhanced brain CT, CT cervical myelography, cervical and intracranial CT angiography and lumbar spine CT both CTDIvol and DLP were lowered by up to 10.9 % (p < 0.001), 17.9 % (p = 0.005), 20.9 % (p < 0.001), and 21.7 % (p = 0.001), respectively, by using ASIR compared with FBP alone. Image quality and noise were similar for both FBP and ASIR.
CONCLUSION: We recommend routine use of iterative reconstruction for neuroradiology CT examinations because this approach affords a significant dose reduction while preserving image quality.
Pubmed
Web of science
Création de la notice
11/04/2014 18:42
Dernière modification de la notice
09/09/2021 7:13
Données d'usage