Characterisation of a commercial flatbed scanner for external beam dose verification using EBT gafchromic and EDR2 films

Details

Serval ID
serval:BIB_98C4C0DA2F90
Type
Inproceedings: an article in a conference proceedings.
Publication sub-type
Abstract (Abstract): shot summary in a article that contain essentials elements presented during a scientific conference, lecture or from a poster.
Collection
Publications
Institution
Title
Characterisation of a commercial flatbed scanner for external beam dose verification using EBT gafchromic and EDR2 films
Title of the conference
11th Annual Meeting of the Scientific Association of Swiss Radiation Oncology (SASRO)
Author(s)
Pachoud M., Hejira N., Bulling S., Do H.P., Bochud F., Moeckli R.
Address
Aarau, Switzerland, March 15-17, 2007
ISBN
0179-7158
Publication state
Published
Issued date
2007
Peer-reviewed
Oui
Volume
183
Series
Strahlentherapie und Onkologie
Pages
181-182
Language
english
Notes
Document Type: Meeting Abstract
Abstract
Objective: The goal of this study was to evaluate the performance of an inexpensive Epson
Perfection V700 flatbed scanner for film dosimetry. The scanner was tested with EBT
Gafchromic and Kodak EDR2 films and compared with a Vidar VXR-16 Dosimetry
Pro scanner - the recognized standard for film dosimetry.
Material and Methods: Measurements were made at the CHUV in Lausanne and Clinique de Genolier in
Genolier. Planar dose distributions with doses in the range 0.1 to 3.5 Gy were
delivered to the EBT Gafchromic and EDR2 films in a solid water phantom.
Sensitometric curves were determined for each scanner-film combination and then
IMRT and percentage depth dose measurements were made.
Results: The uniformity of the scanner response is discussed for the two films and both
scanners. Differences in percentage depth dose measurements and sensitometric
curves for the two films and scanners are shown. The limitations of each scanner for
EBT Gafchromic or EDR2 film dosimetry are described.
Conclusion: An inexpensive Epson Perfection V700 flatbed scanner can be used for EBT
Gafchromic film dosimetry, for example for verifying planar dose distributions for
individual patient IMRT QA.