Characterisation of a commercial flatbed scanner for external beam dose verification using EBT gafchromic and EDR2 films

Détails

ID Serval
serval:BIB_98C4C0DA2F90
Type
Actes de conférence (partie): contribution originale à la littérature scientifique, publiée à l'occasion de conférences scientifiques, dans un ouvrage de compte-rendu (proceedings), ou dans l'édition spéciale d'un journal reconnu (conference proceedings).
Sous-type
Abstract (résumé de présentation): article court qui reprend les éléments essentiels présentés à l'occasion d'une conférence scientifique dans un poster ou lors d'une intervention orale.
Collection
Publications
Institution
Titre
Characterisation of a commercial flatbed scanner for external beam dose verification using EBT gafchromic and EDR2 films
Titre de la conférence
11th Annual Meeting of the Scientific Association of Swiss Radiation Oncology (SASRO)
Auteur⸱e⸱s
Pachoud M., Hejira N., Bulling S., Do H.P., Bochud F., Moeckli R.
Adresse
Aarau, Switzerland, March 15-17, 2007
ISBN
0179-7158
Statut éditorial
Publié
Date de publication
2007
Peer-reviewed
Oui
Volume
183
Série
Strahlentherapie und Onkologie
Pages
181-182
Langue
anglais
Notes
Document Type: Meeting Abstract
Résumé
Objective: The goal of this study was to evaluate the performance of an inexpensive Epson
Perfection V700 flatbed scanner for film dosimetry. The scanner was tested with EBT
Gafchromic and Kodak EDR2 films and compared with a Vidar VXR-16 Dosimetry
Pro scanner - the recognized standard for film dosimetry.
Material and Methods: Measurements were made at the CHUV in Lausanne and Clinique de Genolier in
Genolier. Planar dose distributions with doses in the range 0.1 to 3.5 Gy were
delivered to the EBT Gafchromic and EDR2 films in a solid water phantom.
Sensitometric curves were determined for each scanner-film combination and then
IMRT and percentage depth dose measurements were made.
Results: The uniformity of the scanner response is discussed for the two films and both
scanners. Differences in percentage depth dose measurements and sensitometric
curves for the two films and scanners are shown. The limitations of each scanner for
EBT Gafchromic or EDR2 film dosimetry are described.
Conclusion: An inexpensive Epson Perfection V700 flatbed scanner can be used for EBT
Gafchromic film dosimetry, for example for verifying planar dose distributions for
individual patient IMRT QA.