Optimization of CT pulmonary angiography for pulmonary embolism using task-based image quality assessment and diagnostic reference levels: A multicentric study.

Details

Ressource 1Download: 1-s2.0-S1120179724001601-main.pdf (1836.92 [Ko])
State: Public
Version: Final published version
License: CC BY 4.0
Serval ID
serval:BIB_76F5080F5939
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Optimization of CT pulmonary angiography for pulmonary embolism using task-based image quality assessment and diagnostic reference levels: A multicentric study.
Journal
Physica medica
Author(s)
Viry A., Vitzthum V., Monnin P., Bize J., Rotzinger D., Racine D.
ISSN
1724-191X (Electronic)
ISSN-L
1120-1797
Publication state
Published
Issued date
24/04/2024
Peer-reviewed
Oui
Volume
121
Pages
103365
Language
english
Notes
Publication types: Journal Article
Publication Status: aheadofprint
Abstract
To establish size-specific diagnostic reference levels (DRLs) for pulmonary embolism (PE) based on patient CT examinations performed on 74 CT devices. To assess task-based image quality (IQ) for each device and to investigate the variability of dose and IQ across different CTs. To propose a dose/IQ optimization.
1051 CT pulmonary angiography dose data were collected. DRLs were calculated as the 75th percentile of CT dose index (CTDI) for two patient categories based on the thoracic perimeters. IQ was assessed with two thoracic phantom sizes using local acquisition parameters and three other dose levels. The area under the ROC curve (AUC) of a 2 mm low perfused vessel was assessed with a non-prewhitening with eye-filter model observer. The optimal IQ-dose point was mathematically assessed from the relationship between IQ and dose.
The DRLs of CTDI <sub>vol</sub> were 6.4 mGy and 10 mGy for the two patient categories. 75th percentiles of phantom CTDI <sub>vol</sub> were 6.3 mGy and 10 mGy for the two phantom sizes with inter-quartile AUC values of 0.047 and 0.066, respectively. After the optimization, 75th percentiles of phantom CTDI <sub>vol</sub> decreased to 5.9 mGy and 7.55 mGy and the interquartile AUC values were reduced to 0.025 and 0.057 for the two phantom sizes.
DRLs for PE were proposed as a function of patient thoracic perimeters. This study highlights the variability in terms of dose and IQ. An optimization process can be started individually and lead to a harmonization of practice throughout multiple CT sites.
Keywords
Angiography CT, Diagnostic reference levels, Image quality, Model observers, Optimization, Pulmonary embolism
Pubmed
Create date
26/04/2024 8:28
Last modification date
27/04/2024 7:13
Usage data