Variability of radioiodine measurements in the thyroid.

Details

Ressource 1Download: REF.pdf (113.34 [Ko])
State: Public
Version: Final published version
License: Not specified
It was possible to publish this article open access thanks to a Swiss National Licence with the publisher.
Serval ID
serval:BIB_642DC5AE11D1
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Variability of radioiodine measurements in the thyroid.
Journal
Radiation Protection Dosimetry
Author(s)
Damet J., Bochud F.O., Bailat C., Laedermann J.P., Baechler S.
ISSN
1742-3406 (Electronic)
ISSN-L
0144-8420
Publication state
Published
Issued date
2011
Volume
144
Number
1-4
Pages
326-329
Language
english
Abstract
Monte Carlo simulations were carried out to study the response of a thyroid monitor for measuring intake activities of (125)I and (131)I. The aim of the study was 3-fold: to cross-validate the Monte Carlo simulation programs, to study the response of the detector using different phantoms and to study the effects of anatomical variations. Simulations were performed using the Swiss reference phantom and several voxelised phantoms. Determining the position of the thyroid is crucial for an accurate determination of radiological risks. The detector response using the Swiss reference phantom was in fairly good agreement with the response obtained using adult voxelised phantoms for (131)I, but should be revised for a better calibration for (125)I and for any measurements taken on paediatric patients.
Pubmed
Web of science
Open Access
Yes
Create date
26/04/2011 14:31
Last modification date
14/02/2022 8:55
Usage data