Variability of radioiodine measurements in the thyroid.

Détails

Ressource 1Télécharger: REF.pdf (113.34 [Ko])
Etat: Public
Version: Final published version
Licence: Non spécifiée
It was possible to publish this article open access thanks to a Swiss National Licence with the publisher.
ID Serval
serval:BIB_642DC5AE11D1
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Variability of radioiodine measurements in the thyroid.
Périodique
Radiation Protection Dosimetry
Auteur⸱e⸱s
Damet J., Bochud F.O., Bailat C., Laedermann J.P., Baechler S.
ISSN
1742-3406 (Electronic)
ISSN-L
0144-8420
Statut éditorial
Publié
Date de publication
2011
Volume
144
Numéro
1-4
Pages
326-329
Langue
anglais
Résumé
Monte Carlo simulations were carried out to study the response of a thyroid monitor for measuring intake activities of (125)I and (131)I. The aim of the study was 3-fold: to cross-validate the Monte Carlo simulation programs, to study the response of the detector using different phantoms and to study the effects of anatomical variations. Simulations were performed using the Swiss reference phantom and several voxelised phantoms. Determining the position of the thyroid is crucial for an accurate determination of radiological risks. The detector response using the Swiss reference phantom was in fairly good agreement with the response obtained using adult voxelised phantoms for (131)I, but should be revised for a better calibration for (125)I and for any measurements taken on paediatric patients.
Pubmed
Web of science
Open Access
Oui
Création de la notice
26/04/2011 14:31
Dernière modification de la notice
14/02/2022 8:55
Données d'usage