Right coronary artery flow velocity and volume assessment with spiral K-space sampled breathhold velocity-encoded MRI at 3 tesla: accuracy and reproducibility.

Details

Ressource 1Download: BIB_5FBA70E5350F.P001.pdf (1295.14 [Ko])
State: Public
Version: author
Serval ID
serval:BIB_5FBA70E5350F
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Right coronary artery flow velocity and volume assessment with spiral K-space sampled breathhold velocity-encoded MRI at 3 tesla: accuracy and reproducibility.
Journal
Journal of Magnetic Resonance Imaging
Author(s)
Brandts Anne, Roes Stijntje D., Doornbos Joost, Weiss Robert G., de Roos Albert, Stuber Matthias, Westenberg Jos J. M.
ISSN
1522-2586[electronic], 1053-1807[linking]
Publication state
Published
Issued date
2010
Peer-reviewed
Oui
Volume
31
Number
5
Pages
1215-1223
Language
english
Abstract
PURPOSE: To evaluate accuracy and reproducibility of flow velocity and volume measurements in a phantom and in human coronary arteries using breathhold velocity-encoded (VE) MRI with spiral k-space sampling at 3 Tesla. MATERIALS AND METHODS: Flow velocity assessment was performed using VE MRI with spiral k-space sampling. Accuracy of VE MRI was tested in vitro at five constant flow rates. Reproducibility was investigated in 19 healthy subjects (mean age 25.4 +/- 1.2 years, 11 men) by repeated acquisition in the right coronary artery (RCA). RESULTS: MRI-measured flow rates correlated strongly with volumetric collection (Pearson correlation r = 0.99; P < 0.01). Due to limited sample resolution, VE MRI overestimated the flow rate by 47% on average when nonconstricted region-of-interest segmentation was used. Using constricted region-of-interest segmentation with lumen size equal to ground-truth luminal size, less than 13% error in flow rate was found. In vivo RCA flow velocity assessment was successful in 82% of the applied studies. High interscan, intra- and inter-observer agreement was found for almost all indices describing coronary flow velocity. Reproducibility for repeated acquisitions varied by less than 16% for peak velocity values and by less than 24% for flow volumes. CONCLUSION: 3T breathhold VE MRI with spiral k-space sampling enables accurate and reproducible assessment of RCA flow velocity.
Keywords
Accuracy, Reproducibility, MRI, Coronary Flow Velocity, Phase-Contrast Flow, Signal-To-Noise, Magnetic-Resonance Angiography, Blood-Flow, Quantification, Validation, Excitation, Resolution, Vessels, Ratios