Design and validation of a dosimetric comparison scheme tailored for ultra-high dose-rate electron beams to support multicenter FLASH preclinical studies.

Details

Ressource 1Download: 1-s2.0-S016781402204244X-main.pdf (884.44 [Ko])
State: Public
Version: Final published version
License: CC BY 4.0
Serval ID
serval:BIB_FC59EAAA1608
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Design and validation of a dosimetric comparison scheme tailored for ultra-high dose-rate electron beams to support multicenter FLASH preclinical studies.
Journal
Radiotherapy and oncology
Author(s)
Jorge P.G., Melemenidis S., Grilj V., Buchillier T., Manjappa R., Viswanathan V., Gondré M., Vozenin M.C., Germond J.F., Bochud F., Moeckli R., Limoli C., Skinner L., No H.J., Wu Y.F., Surucu M., Yu A.S., Lau B., Wang J., Schüler E., Bush K., Graves E.E., Maxim P.G., Loo B.W., Bailat C.
ISSN
1879-0887 (Electronic)
ISSN-L
0167-8140
Publication state
Published
Issued date
10/2022
Peer-reviewed
Oui
Volume
175
Pages
203-209
Language
english
Notes
Publication types: Multicenter Study ; Journal Article ; Research Support, Non-U.S. Gov't ; Research Support, N.I.H., Extramural
Publication Status: ppublish
Abstract
We describe a multicenter cross validation of ultra-high dose rate (UHDR) (>= 40 Gy/s) irradiation in order to bring a dosimetric consensus in absorbed dose to water. UHDR refers to dose rates over 100-1000 times those of conventional clinical beams. UHDR irradiations have been a topic of intense investigation as they have been reported to induce the FLASH effect in which normal tissues exhibit reduced toxicity relative to conventional dose rates. The need to establish optimal beam parameters capable of achieving the in vivo FLASH effect has become paramount. It is therefore necessary to validate and replicate dosimetry across multiple sites conducting UHDR studies with distinct beam configurations and experimental set-ups.
Using a custom cuboid phantom with a cylindrical cavity (5 mm diameter by 10.4 mm length) designed to contain three type of dosimeters (thermoluminescent dosimeters (TLDs), alanine pellets, and Gafchromic films), irradiations were conducted at expected doses of 7.5 to 16 Gy delivered at UHDR or conventional dose rates using various electron beams at the Radiation Oncology Departments of the CHUV in Lausanne, Switzerland and Stanford University, CA.
Data obtained between replicate experiments for all dosimeters were in excellent agreement (±3%). In general, films and TLDs were in closer agreement with each other, while alanine provided the closest match between the expected and measured dose, with certain caveats related to absolute reference dose.
In conclusion, successful cross-validation of different electron beams operating under different energies and configurations lays the foundation for establishing dosimetric consensus for UHDR irradiation studies, and, if widely implemented, decrease uncertainty between different sites investigating the mechanistic basis of the FLASH effect.
Keywords
Humans, Electrons, Radiometry, Phantoms, Imaging, Water, Alanine, Dosimetry, FLASH, Intercomparison, Passive dosimeters, UHDR
Pubmed
Web of science
Open Access
Yes
Create date
05/09/2022 9:47
Last modification date
02/02/2023 8:16
Usage data