PORTUGUESE STUDY OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY AND COMPARISON WITH EUROPEAN REFERENCES.
Details
Serval ID
serval:BIB_F2C548B79E1A
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
PORTUGUESE STUDY OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY AND COMPARISON WITH EUROPEAN REFERENCES.
Journal
Radiation protection dosimetry
ISSN
1742-3406 (Electronic)
ISSN-L
0144-8420
Publication state
Published
Issued date
01/06/2018
Peer-reviewed
Oui
Volume
179
Number
4
Pages
391-399
Language
english
Notes
Publication types: Comparative Study ; Journal Article
Publication Status: ppublish
Publication Status: ppublish
Abstract
To characterise the mean glandular dose (MGD) in a sample of healthcare providers for digital mammography in Portugal. To compare the achieved values with European references. The MGD was measured on a poly-methyl-methacrylate phantom (45 mm) for each system using dosimeters. In addition, MGD was estimated using exposure settings collected from mammography exams in clinical context. Data were collected from 25 computed-radiography systems (CR) and 13 integrated digital (DR). For both measurements (phantom and clinical exposures), the average MGD for CR was higher compared to the DR. For CR the mean MGD was 1.85 mGy (CC projection) and 2.10 mGy (MLO projection). For DR systems the corresponding values were 1.54 mGy (CC) and 1.68 mGy (MLO). The average MGD obtained using both methods and for both technologies is within the acceptable reference range proposed by European guidelines (<2.5 mGy). Dose Reference Levels implementation should be the next step to optimise mammography practice in Portugal.
Keywords
Adult, Aged, Breast/radiation effects, Europe, Female, Humans, Mammography, Middle Aged, Phantoms, Imaging, Portugal, Radiation Dosage, Reference Values