New, Virtually Wall-less Cannulas Designed for Augmented Venous Drainage in Minimally Invasive Cardiac Surgery.

Details

Serval ID
serval:BIB_F21312A0E0F5
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
New, Virtually Wall-less Cannulas Designed for Augmented Venous Drainage in Minimally Invasive Cardiac Surgery.
Journal
Innovations (Philadelphia, Pa.)
Author(s)
von Segesser L.K., Berdajs D., Abdel-Sayed S., Tozzi P., Ferrari E., Maisano F.
ISSN
1559-0879 (Electronic)
ISSN-L
1556-9845
Publication state
Published
Issued date
2016
Peer-reviewed
Oui
Volume
11
Number
4
Pages
278-281
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Abstract
Inadequate venous drainage during minimally invasive cardiac surgery becomes most evident when the blood trapped in the pulmonary circulation floods the surgical field. The present study was designed to assess the in vivo performance of new, thinner, virtually wall-less, venous cannulas designed for augmented venous drainage in comparison to traditional thin-wall cannulas.
Remote cannulation was realized in 5 bovine experiments (74.0 ± 2.4 kg) with percutaneous venous access over the wire, serial dilation up to 18 F and insertion of either traditional 19 F thin wall, wire-wound cannulas, or through the same access channel, new, thinner, virtually wall-less, braided cannulas designed for augmented venous drainage. A standard minimal extracorporeal circuit set with a centrifugal pump and a hollow fiber membrane oxygenator, but no in-line reservoir was used. One hundred fifty pairs of pump-flow and required pump inlet pressure values were recorded with calibrated pressure transducers and a flowmeter calibrated by a volumetric tank and timer at increasing pump speed from 1500 RPM to 3500 RPM (500-RPM increments).
Pump flow accounted for 1.73 ± 0.85 l/min for wall-less versus 1.17 ± 0.45 l/min for thin wall at 1500 RPM, 3.91 ± 0.86 versus 3.23 ± 0.66 at 2500 RPM, 5.82 ± 1.05 versus 4.96 ± 0.81 at 3500 RPM. Pump inlet pressure accounted for 9.6 ± 9.7 mm Hg versus 4.2 ± 18.8 mm Hg for 1500 RPM, -42.4 ± 26.7 versus -123 ± 51.1 at 2500 RPM, and -126.7 ± 55.3 versus -313 ± 116.7 for 3500 RPM.
At the well-accepted pump inlet pressure of -80 mm Hg, the new, thinner, virtually wall-less, braided cannulas provide unmatched venous drainage in vivo. Early clinical analyses have confirmed these findings.

Keywords
Animals, Cardiac Surgical Procedures/instrumentation, Catheterization, Cattle, Drainage/instrumentation, Drainage/methods, Equipment Design, Minimally Invasive Surgical Procedures/instrumentation, Minimally Invasive Surgical Procedures/methods, Models, Animal, Veins
Pubmed
Create date
02/09/2016 10:23
Last modification date
20/08/2019 17:19
Usage data