Long-term outcomes with biodegradable polymer sirolimus-eluting stents versus durable polymer everolimus-eluting stents in ST-segment elevation myocardial infarction: 5-year follow-up of the BIOSTEMI randomised superiority trial.

Details

Serval ID
serval:BIB_EEABEFE0006A
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Long-term outcomes with biodegradable polymer sirolimus-eluting stents versus durable polymer everolimus-eluting stents in ST-segment elevation myocardial infarction: 5-year follow-up of the BIOSTEMI randomised superiority trial.
Journal
Lancet
Author(s)
Iglesias J.F., Roffi M., Losdat S., Muller O., Degrauwe S., Kurz D.J., Haegeli L., Weilenmann D., Kaiser C., Tapponnier M., Cook S., Cuculi F., Heg D., Windecker S., Pilgrim T.
ISSN
1474-547X (Electronic)
ISSN-L
0140-6736
Publication state
Published
Issued date
25/11/2023
Peer-reviewed
Oui
Volume
402
Number
10416
Pages
1979-1990
Language
english
Notes
Publication types: Randomized Controlled Trial ; Journal Article
Publication Status: ppublish
Abstract
Biodegradable polymer sirolimus-eluting stents improve early stent-related clinical outcomes compared to durable polymer everolimus-eluting stents in patients with ST-segment elevation myocardial infarction (STEMI) undergoing primary percutaneous coronary intervention. The long-term advantages of biodegradable polymer sirolimus-eluting stents after complete degradation of its polymer coating in patients with STEMI remains however uncertain.
BIOSTEMI Extended Survival (BIOSTEMI ES) was an investigator-initiated, follow-up extension study of the BIOSTEMI prospective, multicentre, single-blind, randomised superiority trial that compared biodegradable polymer sirolimus-eluting stents with durable polymer everolimus-eluting stents in patients with STEMI undergoing primary percutaneous coronary intervention at ten hospitals in Switzerland. All individuals who had provided written informed consent for participation in the BIOSTEMI trial were eligible for this follow-up study. The primary endpoint was target lesion failure, defined as a composite of cardiac death, target vessel myocardial re-infarction, or clinically indicated target lesion revascularisation, at 5 years. Superiority of biodegradable polymer sirolimus-eluting stents over durable polymer everolimus-eluting stents was declared if the Bayesian posterior probability for a rate ratio (RR) of less than 1 was greater than 0·975. Analyses were performed according to the intention-to-treat principle. The study was registered with ClinicalTrials.gov, NCT05484310.
Between April 26, 2016, and March 9, 2018, 1300 patients with STEMI (1622 lesions) were randomly allocated in a 1:1 ratio to treatment with biodegradable polymer sirolimus-eluting stents (649 patients, 816 lesions) or durable polymer everolimus-eluting stents (651 patients, 806 lesions). At 5 years, the primary composite endpoint of target lesion failure occurred in 50 (8%) patients treated with biodegradable polymer sirolimus-eluting stents and in 72 (11%) patients treated with durable polymer everolimus-eluting stents (difference of -3%; RR 0·70, 95% Bayesian credible interval 0·51-0·95; Bayesian posterior probability for superiority 0·988).
In patients undergoing primary percutaneous coronary intervention for STEMI, biodegradable polymer sirolimus-eluting stents were superior to durable polymer everolimus-eluting stents with respect to target lesion failure at 5 years of follow-up. The difference was driven by a numerically lower risk for ischaemia-driven target lesion revascularisation.
Biotronik.
Keywords
Humans, Sirolimus/therapeutic use, Everolimus/therapeutic use, Follow-Up Studies, ST Elevation Myocardial Infarction/surgery, ST Elevation Myocardial Infarction/drug therapy, Drug-Eluting Stents, Polymers, Bayes Theorem, Single-Blind Method, Prospective Studies, Treatment Outcome, Absorbable Implants, Myocardial Infarction/etiology, Percutaneous Coronary Intervention/methods