Limitations using the vacuum-assist venous drainage technique during cardiopulmonary bypass procedures

Details

Serval ID
serval:BIB_ED63B252BF7C
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Limitations using the vacuum-assist venous drainage technique during cardiopulmonary bypass procedures
Journal
Journal of Extra-Corporeal Technology
Author(s)
Jegger  D., Tevaearai  H. T., Mueller  X. M., Horisberger  J., von Segesser  L. K.
ISSN
0022-1058 (Print)
Publication state
Published
Issued date
09/2003
Volume
35
Number
3
Pages
207-11
Notes
In Vitro
Journal Article --- Old month value: Sep
Abstract
Vacuum-assist venous drainage (VAVD) can increase venous blood return during cardiopulmonary bypass (CPB) procedures. However, the negative pressure created in the closed cardiotomy reservoir can be transmitted to the oxygenator if a nonocclusive or centrifugal arterial pump is used, resulting in bubble transgression (BT) from the gas to blood compartment of the oxygenator. We analyzed the vacuum pressure required to produce BT using an in vitro circuit including successively a closed reservoir, a pump (centrifugal or roller), and an oxygenator. A constant hydrostatic pressure was maintained onto the oxygenator. Vacuum was applied on the cardiotomy reservoir, progressively increasing negative pressure from 0 to -80 mmHg and monitoring BT with a bubble detector. Six different oxygenators were compared. A partially occlusive roller pump and a centrifugal pump were compared to a control, which was without any pump. A mean negative pressure of -53 +/- 7 mmHg was necessary to produce BT in all the oxygenators in the absence of a pump. The presence of a centrifugal pump between the reservoir and the oxygenator significantly increased the negative pressure required to produce BT compared to the control (-67 +/- 7 mmHg, p < .05). No bubbles were detected using the roller pump (> -80 mmHg needed for BT), thus statistically significant when compared to the centrifugal pump (p < .05). The centrifugal pump offers significant resistance to BT but not as much compared to the roller pump, though BT cannot be prevented if the pump is turned off while the vacuum remains on the reservoir. Therefore, VAVD is a safe technique as long as the perfusionist stops the vacuum when the arterial pump is no longer in use.
Keywords
Adult Cardiopulmonary Bypass/*instrumentation/methods Extracorporeal Membrane Oxygenation/*instrumentation Humans *Oxygenators/standards Suction/*instrumentation Switzerland
Pubmed
Create date
28/01/2008 10:28
Last modification date
20/08/2019 17:15
Usage data