Pulse oximetry for hypoxemia: a warning to users and manufacturers.
Details
Serval ID
serval:BIB_E031E78C57B6
Type
Article: article from journal or magazin.
Publication sub-type
Review (review): journal as complete as possible of one specific subject, written based on exhaustive analyses from published work.
Collection
Publications
Institution
Title
Pulse oximetry for hypoxemia: a warning to users and manufacturers.
Journal
Intensive care medicine
ISSN
0342-4642
Publication state
Published
Issued date
1989
Peer-reviewed
Oui
Volume
15
Number
8
Pages
540-2
Language
english
Notes
Publication types: Journal Article ; Review - Publication Status: ppublish
Abstract
Pulse oximetry represents a major advance in patient monitoring, but measurement below 70 to 80% saturation has important limitations. Several authors have tested pulse oximetry at low saturations with conflicting results. A review of these data indicates that every patient with a pulse oximeter value below 75 to 80% SaO2 should have one or more invasive measurements of the arterial SaO2 in order to avoid undetected severe hypoxemia.
Keywords
Animals, Anoxia, Child, Critical Care, Dogs, False Negative Reactions, Humans, Infant, Monitoring, Physiologic, Oximetry
Pubmed
Web of science
Create date
25/01/2008 10:06
Last modification date
20/08/2019 16:04