The Surgical Approach to the Anterior Nucleus of Thalamus in Patients With Refractory Epilepsy: Experience from the International Multicenter Registry (MORE).
Details
Serval ID
serval:BIB_D34F0284B373
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
The Surgical Approach to the Anterior Nucleus of Thalamus in Patients With Refractory Epilepsy: Experience from the International Multicenter Registry (MORE).
Journal
Neurosurgery
Working group(s)
MORE investigators
ISSN
1524-4040 (Electronic)
ISSN-L
0148-396X
Publication state
Published
Issued date
01/01/2019
Peer-reviewed
Oui
Volume
84
Number
1
Pages
141-150
Language
english
Notes
Publication types: Journal Article ; Observational Study
Publication Status: ppublish
Publication Status: ppublish
Abstract
The Medtronic Registry for Epilepsy (MORE; Medtronic Inc, Dublin, Ireland) is an open label observational study evaluating the long-term effectiveness, safety, and performance of deep brain stimulation (DBS) of the anterior nucleus of thalamus (ANT) for the treatment of refractory epilepsy.
To compare the difference in success rate of placing contacts at ANT-target region (ANT-TR) between transventricular (TV) and extraventricular (EV) lead trajectories in 73 ANT-DBS implants in 17 European centers participating in the MORE registry.
The success rate of placing contacts at ANT-TR was evaluated using a screening method combining both individual patient imaging information and stereotactic atlas information to identify contacts at ANT-TR.
EV lead trajectory was used in 53% of the trajectories. Approximately, 90% of the TV lead trajectories had at least 1 contact at ANT-TR, vs only 71% of the EV lead trajectories. The success rate for placing at least 1 contact at ANT-TR bilaterally was 84% for TV implants and 58% for EV implants (P < .05; Fisher's exact). No intracranial bleedings were observed, but 1 cortical infarct was reported following EV lead trajectory.
The results of this registry support the use of TV lead trajectories for ANT-DBS as they have a higher probability in placing contacts at ANT-TR, without appearing to compromise procedural safety. Follow-up data collection is continuing in the MORE registry. These data will provide outcomes associated with TV and EV trajectories.
To compare the difference in success rate of placing contacts at ANT-target region (ANT-TR) between transventricular (TV) and extraventricular (EV) lead trajectories in 73 ANT-DBS implants in 17 European centers participating in the MORE registry.
The success rate of placing contacts at ANT-TR was evaluated using a screening method combining both individual patient imaging information and stereotactic atlas information to identify contacts at ANT-TR.
EV lead trajectory was used in 53% of the trajectories. Approximately, 90% of the TV lead trajectories had at least 1 contact at ANT-TR, vs only 71% of the EV lead trajectories. The success rate for placing at least 1 contact at ANT-TR bilaterally was 84% for TV implants and 58% for EV implants (P < .05; Fisher's exact). No intracranial bleedings were observed, but 1 cortical infarct was reported following EV lead trajectory.
The results of this registry support the use of TV lead trajectories for ANT-DBS as they have a higher probability in placing contacts at ANT-TR, without appearing to compromise procedural safety. Follow-up data collection is continuing in the MORE registry. These data will provide outcomes associated with TV and EV trajectories.
Keywords
Anterior Thalamic Nuclei/surgery, Deep Brain Stimulation/instrumentation, Deep Brain Stimulation/methods, Drug Resistant Epilepsy/surgery, Humans, Implantable Neurostimulators, Registries