Generic sample preparation and dual polarity liquid chromatography-time-of-flight mass spectrometry for high-throughput screening in doping analysis.
Details
Serval ID
serval:BIB_D16AB8C6B02E
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Generic sample preparation and dual polarity liquid chromatography-time-of-flight mass spectrometry for high-throughput screening in doping analysis.
Journal
Drug testing and analysis
ISSN
1942-7611 (Electronic)
ISSN-L
1942-7603
Publication state
Published
Issued date
06/2009
Peer-reviewed
Oui
Volume
1
Number
6
Pages
250-266
Language
english
Notes
Publication types: Journal Article ; Validation Studies
Publication Status: ppublish
Publication Status: ppublish
Abstract
The requirements on initial testing in doping control are getting tighter regarding efficiency and speed while the scope of analytes is getting more diverse and, consequently, the need for high-throughput methods is apparent. In this study, a comprehensive screening method for doping agents in human urine is presented, based on solid phase extraction (SPE) and liquid chromatography-time-of-flight mass spectrometry (LCTOFMS). The method covers most of the compound groups in the list of prohibited substances by World Anti-Doping Agency (WADA). Mixed-mode SPE on two types of sorbent and the use of negative ionization mode besides the commonly used positive mode in electrospray ionization (ESI) allowed detection of acidic compounds, such as sulpho-conjugated metabolites. A run time of 8 minutes for each of the two ESI polarities was achieved. The method was validated regarding relative ionization efficiency, selectivity and signal to noise at the WADA's minimum required performance limit (MRPL) level, resulting in the acceptance of 197 compounds. A selection of 20 compounds was submitted for a more thorough validation, including extraction recovery, repeatability and linearity. Recovery and linearity (R(2)) varied mainly between 83-115% and 0.78-0.99, respectively. Median values for repeatability at the MRPL and 10 x MRPL levels were below 20%. A mean and median mass accuracy of 1.2 and 0.80 mDa, respectively, was achieved. The present method represents at the moment the widest coverage of low molecular weight prohibited substances for the screening in sports, providing an approach for further rationalisation of the analytical work-flow in the doping control laboratories.
Keywords
Chromatography, Liquid/methods, Doping in Sports, High-Throughput Screening Assays/methods, Humans, Molecular Weight, Reproducibility of Results, Solid Phase Extraction/methods, Spectrometry, Mass, Electrospray Ionization/methods, Substance Abuse Detection/methods
Pubmed
Web of science
Create date
02/05/2017 14:27
Last modification date
20/08/2019 15:51