Techniques and standards in intraoperative graft verification by transit time flow measurement after coronary artery bypass graft surgery: a critical review.

Details

Serval ID
serval:BIB_BC21785BAA14
Type
Article: article from journal or magazin.
Publication sub-type
Review (review): journal as complete as possible of one specific subject, written based on exhaustive analyses from published work.
Collection
Publications
Institution
Title
Techniques and standards in intraoperative graft verification by transit time flow measurement after coronary artery bypass graft surgery: a critical review.
Journal
European journal of cardio-thoracic surgery
Author(s)
Niclauss L.
ISSN
1873-734X (Electronic)
ISSN-L
1010-7940
Publication state
Published
Issued date
01/2017
Peer-reviewed
Oui
Volume
51
Number
1
Pages
26-33
Language
english
Notes
Publication types: Journal Article ; Review
Publication Status: ppublish
Abstract
Transit time flow measurement (TTFM) is a quality control tool for intraoperative graft evaluation in coronary artery bypass graft (CABG) surgery. A critical review of the literature available using TTFM in CABG surgery is the focus of this article. The main objectives will be to detail precise parameters for flow evaluation, to show limitations of TTFM and to prove its predictive impact on postoperative graft failure rate. Publications listed in the PubMed database were reviewed, searching for intraoperative graft verification in coronary surgery by TTFM, with postoperative imaging follow-up (FU) modality and with a special focus on publications released after European guidelines from 2010. Nine included publications revealed an overall graft failure rate of ∼12%. Mean graft flow had a positive predictive value in the largest study, and cut-offs, of at least 20 ml/min for internal mammary artery (IMA) grafts, therein partially confirming guidelines, and 30-40 ml/min for saphenous venous grafts (SVGs) were proposed. An explicit correlation between graft flow, patency rate and severity of coronary stenosis, by indicating the fractional flow reserve, was found for IMA grafts. Increased pulsatility index and increased systolic reverse flow probably predict worse outcome and may help identifying competitive flow. Diastolic filling, rarely indicated, could not be confirmed as the predictive marker. No significant correlation of TTFM and graft failure rate for radial and other arterial grafts could be found, partially due to the small number of these types of grafts analysed. Larger target vessels and lower postoperative CK-MB levels may predict better graft patency rates. Low sensitivity for TTFM to reliably detect graft failure is certainly a major issue, as found in randomized analyses. However, methodical limitations and varying threshold values for TTFM render a general consensus difficult. Influence of quantity (vessel territory distribution) and quality (myocardial scar) of the graft perfusion area, on TTFM and FU outcome, was not included by anyone and should be part of future research. TTFM is probably not the tool of choice to detect progressive late graft failure of SVG. Peroperative TTFM values should be correlated with one type of conduit, differentiating between early and late graft failure (by applying a uniform, appropriated definition), to precise and confirm threshold values.

Keywords
Blood Flow Velocity/physiology, Coronary Angiography, Coronary Artery Bypass/methods, Coronary Artery Disease/diagnosis, Coronary Artery Disease/surgery, Coronary Circulation/physiology, Graft Occlusion, Vascular/diagnosis, Graft Occlusion, Vascular/physiopathology, Humans, Intraoperative Period, Mammary Arteries/diagnostic imaging, Mammary Arteries/physiopathology, Mammary Arteries/transplantation, Postoperative Period, Vascular Patency, Coronary surgery, Quality control, Transit time flow measurement
Pubmed
Web of science
Open Access
Yes
Create date
18/06/2016 16:07
Last modification date
20/08/2019 15:30
Usage data