Optimization of contrast-enhanced spectral mammography depending on clinical indication

Details

Serval ID
serval:BIB_AF26257FE8E2
Type
Article: article from journal or magazin.
Collection
Publications
Title
Optimization of contrast-enhanced spectral mammography depending on clinical indication
Journal
J Med Imaging (Bellingham)
Author(s)
Dromain C., Canale S., Saab-Puong S., Carton A. K., Muller S., Fallenberg E. M.
ISSN-L
2329-4302 (Print)2329-4302 (Linking)
Publication state
Published
Issued date
2014
Peer-reviewed
Oui
Volume
1
Number
3
Pages
033506
Language
english
Notes
Dromain, ClarisseCanale, SandraSaab-Puong, SylvieCarton, Ann-KatherineMuller, SergeFallenberg, Eva MariaengBellingham, Wash.2015/07/15 06:00J Med Imaging (Bellingham). 2014 Oct;1(3):033506. doi: 10.1117/1.JMI.1.3.033506. Epub 2014 Dec 30.
Abstract
The objective is to optimize low-energy (LE) and high-energy (HE) exposure parameters of contrast-enhanced spectral mammography (CESM) examinations in four different clinical applications for which different levels of average glandular dose (AGD) and ratios between LE and total doses are required. The optimization was performed on a Senographe DS with a SenoBright(R) upgrade. Simulations were performed to find the optima by maximizing the contrast-to-noise ratio (CNR) on the recombined CESM image using different targeted doses and LE image quality. The linearity between iodine concentration and CNR as well as the minimal detectable iodine concentration was assessed. The image quality of the LE image was assessed on the CDMAM contrast-detail phantom. Experiments confirmed the optima found on simulation. The CNR was higher for each clinical indication than for SenoBright(R), including the screening indication for which the total AGD was 22% lower. Minimal iodine concentrations detectable in the case of a 3-mm-diameter round tumor were 12.5% lower than those obtained for the same dose in the clinical routine. LE image quality satisfied EUREF acceptable limits for threshold contrast. This newly optimized set of acquisition parameters allows increased contrast detectability compared to parameters currently used without a significant loss in LE image quality.
Create date
16/09/2016 11:14
Last modification date
20/08/2019 16:18
Usage data