The value of <sup>18</sup>F-FDG PET/CT for the diagnosis of device-related infections in patients with a left ventricular assist device: a systematic review and meta-analysis.

Details

Ressource 1Download: 32594196_BIB_A3BF5E76C9BD.pdf (965.23 [Ko])
State: Public
Version: Final published version
License: CC BY 4.0
Serval ID
serval:BIB_A3BF5E76C9BD
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
The value of <sup>18</sup>F-FDG PET/CT for the diagnosis of device-related infections in patients with a left ventricular assist device: a systematic review and meta-analysis.
Journal
European journal of nuclear medicine and molecular imaging
Author(s)
Ten Hove D., Treglia G., Slart RHJA, Damman K., Wouthuyzen-Bakker M., Postma D.F., Gheysens O., Borra RJH, Mecozzi G., van Geel P.P., Sinha B., Glaudemans AWJM
ISSN
1619-7089 (Electronic)
ISSN-L
1619-7070
Publication state
Published
Issued date
01/2021
Peer-reviewed
Oui
Volume
48
Number
1
Pages
241-253
Language
english
Notes
Publication types: Journal Article ; Meta-Analysis ; Research Support, Non-U.S. Gov't ; Review ; Systematic Review
Publication Status: ppublish
Abstract
Left ventricular assist devices (LVADs) are increasingly used for the treatment of advanced heart failure. LVADs improve quality of life and decrease mortality, but the driveline carries substantial risk for major infections. These device-related LVAD and driveline infections are difficult to diagnose with conventional imaging. We reviewed and analysed the current literature on the additive value of <sup>18</sup> F-fluorodeoxyglucose positron emission tomography combined with computed tomography (FDG-PET/CT) imaging for the diagnosis of LVAD-related infections."
We performed a systematic literature review using several databases from their inception until the 31st of December, 2019. Studies investigating the diagnostic performance of FDG-PET/CT in patients with suspected LVAD infection were retrieved. After a bias risk assessment using QUADAS-2, a study-aggregate meta-analysis was performed on a per examination-based analysis.
A total of 10 studies were included in the systematic review, eight of which were also eligible for study-aggregate meta-analysis. For the meta-analysis, a total of 256 FDG-PET/CT scans, examining pump/pocket and/or driveline infection, were acquired in 230 patients. Pooled sensitivity of FDG-PET/CT was 0.95 (95% confidence interval (CI) 0.89-0.97) and pooled specificity was 0.91 (95% CI 0.54-0.99) for the diagnosis of device-related infection. For pump/pocket infection, sensitivity and specificity of FDG-PET/CT were 0.97 (95%CI 0.69-1.00) and 0.93 (95%CI 0.64-0.99), respectively. For driveline infection, sensitivity and specificity were 0.96 (95%CI 0.88-0.99) and 0.99 (95%CI 0.13-1.00) respectively. Significant heterogeneity existed across studies for specificity, mostly caused by differences in scan procedures. Predefined criteria for suspicion of LVAD and/or driveline infection were lacking in all included studies.
FDG-PET/CT is a valuable tool for assessment of device-related infection in LVAD patients, with high sensitivity and high, albeit variable, specificity. Standardization of FDG-PET/CT procedures and criteria for suspected device-related LVAD infections are needed for consistent reporting of FDG-PET/CT scans.
Keywords
Fluorodeoxyglucose F18, Heart-Assist Devices/adverse effects, Humans, Positron Emission Tomography Computed Tomography, Positron-Emission Tomography, Prosthesis-Related Infections/diagnostic imaging, Quality of Life, Radiopharmaceuticals, Sensitivity and Specificity, 18F-FDG PET/CT, LVAD infection, Meta-analysis, Systematic review