A mathematical model for the evaluation of the behaviour during flexion of condylar-type knee prostheses

Details

Serval ID
serval:BIB_992AB2BC4A8E
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
A mathematical model for the evaluation of the behaviour during flexion of condylar-type knee prostheses
Journal
Journal of Biomechanics
Author(s)
Essinger  J. R., Leyvraz  P. F., Heegard  J. H., Robertson  D. D.
ISSN
0021-9290 (Print)
Publication state
Published
Issued date
1989
Volume
22
Number
11-12
Pages
1229-41
Notes
Journal Article
Abstract
A 3D knee model was developed in order to evaluate the mechanical behaviour during flexion of condylar-type knee prosthesis. Based on the total energy minimization principle, it takes into account the articular surfaces (the tibial surface being deformable), the body weight, and the patello femoral joint. It generates the kinematics of the joint, the motion of the centre of contact, the quadriceps forces, the pressure distribution on the tibial plateau, and ligament lengths and forces between 0 and 120 degrees of flexion. The results for ten digitized knees and the commercially available prostheses are presented. They are in general agreement with experimental results published in the literature. It is concluded that this computer program may be, within its limitations, a useful tool in the preliminary evaluation of new condylar-type knee prosthesis designs.
Keywords
Biomechanics Body Weight Humans Knee Joint/*physiology *Knee Prosthesis Ligaments, Articular/*physiology Mathematics Models, Biological