Prevention of esophageal lesions during atrial fibrillation catheter ablation using esophageal temperature monitoring: A systematic review and meta-analysis.

Details

Serval ID
serval:BIB_94ACF000302A
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Prevention of esophageal lesions during atrial fibrillation catheter ablation using esophageal temperature monitoring: A systematic review and meta-analysis.
Journal
Pacing and clinical electrophysiology
Author(s)
Salihu A., Lu H., Maurizi N., Tzimas G., Herrera Siklody C., Le Bloa M., Domenichini G., Teres C., Hugelshofer S., Monney P., Pruvot E., Muller O., Antiochos P., Pascale P.
ISSN
1540-8159 (Electronic)
ISSN-L
0147-8389
Publication state
In Press
Peer-reviewed
Oui
Language
english
Notes
Publication types: Journal Article
Publication Status: aheadofprint
Abstract
The use of esophageal temperature monitoring (ETM) for the prevention of esophageal injury during atrial fibrillation (AF) ablation is often advocated. However, evidence supporting its use is scarce and controversial. We therefore aimed to review the evidence assessing the efficacy of ETM for the prevention of esophageal injury.
We performed a meta-analysis and systematic review of the available literature from inception to December 31, 2022. All studies comparing the use of ETM, versus no ETM, during radiofrequency (RF) AF ablation and which reported the incidence of endoscopically detected esophageal lesions (EDELs) were included.
Eleven studies with a total of 1112 patients undergoing RF AF ablation were identified. Of those patients, 627 were assigned to ETM (56%). The overall incidence of EDELs was 9.8%. The use of ETM during AF ablation was associated with a non significant increase in the incidence of EDELs (12.3% with ETM, vs. 6.6 % without ETM, odds ratio, 1.44, 95%CI, 0.49, 4.22, p = .51, I <sup>2 </sup> = 72%). The use of ETM was associated with a significant increase in the energy delivered specifically on the posterior wall compared to patients without ETM (mean power difference: 5.13 Watts, 95% CI, 1.52, 8.74, p = .005).
The use of ETM does not reduce the incidence of EDELs during RF AF ablation. The higher energy delivered on the posterior wall is likely attributable to a false sense of safety that may explain the lack of benefit of ETM. Further randomized controlled trials are needed to provide conclusive results.
Keywords
ablation, atrial fibrillation, esophageal lesion, radiofrequency, temperature probe