Atrial assist device, a new alternative to lifelong anticoagulation?
Journal
Swiss Medical Weekly
Author(s)
Abdelnour-Berchtold E., Tozzi P., Siniscalchi G., Hayoz D., von Segesser L.K.
ISSN
1424-7860
Publication state
Published
Issued date
2009
Peer-reviewed
Oui
Volume
139
Number
5-6
Pages
82-87
Language
english
Abstract
OBJECTIVE: Atrial fibrillation is a very common heart arrhythmia, associated with a five-fold increase in the risk of embolic strokes. Treatment strategies encompass palliative drugs or surgical procedures all of which can restore sinus rhythm. Unfortunately, atria often fail to recover their mechanical function and patients therefore require lifelong anticoagulation therapy. A motorless volume displacing device (Atripump) based on artificial muscle technology, positioned on the external surface of atrium could avoid the need of oral anticoagulation and its haemorrhagic complications. An animal study was conducted in order to assess the haemodynamic effects that such a pump could provide. METHODS: Atripump is a dome-shape siliconecoated nitinol actuator sewn on the external surface of the atrium. It is driven by a pacemaker-like control unit. Five non-anticoagulated sheep were selected for this experiment. The right atrium was surgically exposed, the device sutured and connected. Haemodynamic parameters and intracardiac ultrasound (ICUS) data were recorded in each animal and under three conditions; baseline; atrial fibrillation (AF); atripump assisted AF (aaAF). RESULTS: In two animals, after 20 min of AF, small thrombi appeared in the right atrial appendix and were washed out once the pump was turned on. Assistance also enhanced atrial ejection fraction. 31% baseline; 5% during AF; 20% under aaAF. Right atrial systolic surfaces (cm2) were; 5.2 +/- 0.3 baseline; 6.2 +/- 0.1 AF; 5.4 +/- 0.3 aaAF. CONCLUSION: This compact and reliable pump seems to restore the atrial "kick" and prevents embolic events. It could avoid long-term anticoagulation therapy and open new hopes in the care of end-stage heart failure.
Keywords
Alloys, Animals, Anticoagulants, Atrial Fibrillation, Atrial Function, Right, Equipment Design, Heart-Assist Devices, Materials Testing, Sheep, Stroke, Thromboembolism
Pubmed
Web of science
Create date
29/05/2009 10:33
Last modification date
20/08/2019 14:52
Usage data