Steady-state equilibrium phase inversion recovery ON-resonant water suppression (IRON) MR angiography in conjunction with superparamagnetic nanoparticles. A robust technique for imaging within a wide range of contrast agent dosages.
Details
Download: BIB_8AA75E905EFF.P001.pdf (7943.32 [Ko])
State: Public
Version: author
State: Public
Version: author
Serval ID
serval:BIB_8AA75E905EFF
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Steady-state equilibrium phase inversion recovery ON-resonant water suppression (IRON) MR angiography in conjunction with superparamagnetic nanoparticles. A robust technique for imaging within a wide range of contrast agent dosages.
Journal
Journal of Magnetic Resonance Imaging
ISSN
1522-2586 (Electronic)
ISSN-L
1053-1807
Publication state
Published
Issued date
2013
Peer-reviewed
Oui
Volume
38
Number
4
Pages
836-844
Language
english
Notes
Publication types: Journal Article ; Research Support, American Recovery and Reinvestment Act ; Research Support, N.I.H., Extramural ; Research Support, Non-U.S. Gov'tPublication Status: ppublish
Abstract
PURPOSE: To investigate the ability of inversion recovery ON-resonant water suppression (IRON) in conjunction with P904 (superparamagnetic nanoparticles which consisting of a maghemite core coated with a low-molecular-weight amino-alcohol derivative of glucose) to perform steady-state equilibrium phase MR angiography (MRA) over a wide dose range.
MATERIALS AND METHODS: Experiments were approved by the institutional animal care committee. Rabbits (n = 12) were imaged at baseline and serially after the administration of 10 incremental dosages of 0.57-5.7 mgFe/Kg P904. Conventional T1-weighted and IRON MRA were obtained on a clinical 1.5 Tesla (T) scanner to image the thoracic and abdominal aorta, and peripheral vessels. Contrast-to-noise ratios (CNR) and vessel sharpness were quantified.
RESULTS: Using IRON MRA, CNR and vessel sharpness progressively increased with incremental dosages of the contrast agent P904, exhibiting constantly higher contrast values than T1 -weighted MRA over a very wide range of contrast agent doses (CNR of 18.8 ± 5.6 for IRON versus 11.1 ± 2.8 for T1 -weighted MRA at 1.71 mgFe/kg, P = 0.02 and 19.8 ± 5.9 for IRON versus -0.8 ± 1.4 for T1-weighted MRA at 3.99 mgFe/kg, P = 0.0002). Similar results were obtained for vessel sharpness in peripheral vessels, (Vessel sharpness of 46.76 ± 6.48% for IRON versus 33.20 ± 3.53% for T1-weighted MRA at 1.71 mgFe/Kg, P = 0.002, and of 48.66 ± 5.50% for IRON versus 19.00 ± 7.41% for T1-weighted MRA at 3.99 mgFe/Kg, P = 0.003).
CONCLUSION: Our study suggests that quantitative CNR and vessel sharpness after the injection of P904 are consistently higher for IRON MRA when compared with conventional T1-weighted MRA. These findings apply for a wide range of contrast agent dosages.
MATERIALS AND METHODS: Experiments were approved by the institutional animal care committee. Rabbits (n = 12) were imaged at baseline and serially after the administration of 10 incremental dosages of 0.57-5.7 mgFe/Kg P904. Conventional T1-weighted and IRON MRA were obtained on a clinical 1.5 Tesla (T) scanner to image the thoracic and abdominal aorta, and peripheral vessels. Contrast-to-noise ratios (CNR) and vessel sharpness were quantified.
RESULTS: Using IRON MRA, CNR and vessel sharpness progressively increased with incremental dosages of the contrast agent P904, exhibiting constantly higher contrast values than T1 -weighted MRA over a very wide range of contrast agent doses (CNR of 18.8 ± 5.6 for IRON versus 11.1 ± 2.8 for T1 -weighted MRA at 1.71 mgFe/kg, P = 0.02 and 19.8 ± 5.9 for IRON versus -0.8 ± 1.4 for T1-weighted MRA at 3.99 mgFe/kg, P = 0.0002). Similar results were obtained for vessel sharpness in peripheral vessels, (Vessel sharpness of 46.76 ± 6.48% for IRON versus 33.20 ± 3.53% for T1-weighted MRA at 1.71 mgFe/Kg, P = 0.002, and of 48.66 ± 5.50% for IRON versus 19.00 ± 7.41% for T1-weighted MRA at 3.99 mgFe/Kg, P = 0.003).
CONCLUSION: Our study suggests that quantitative CNR and vessel sharpness after the injection of P904 are consistently higher for IRON MRA when compared with conventional T1-weighted MRA. These findings apply for a wide range of contrast agent dosages.