Le danger de la radiotherapie en presence d'un stimulateur cardiaque. [Hazards of radiotherapy in the presence of a cardiac pacemaker]

Details

Serval ID
serval:BIB_7B58990D7007
Type
Article: article from journal or magazin.
Publication sub-type
Case report (case report): feedback on an observation with a short commentary.
Collection
Publications
Institution
Title
Le danger de la radiotherapie en presence d'un stimulateur cardiaque. [Hazards of radiotherapy in the presence of a cardiac pacemaker]
Journal
Schweizerische Rundschau fur Medizin Praxis
Author(s)
Jaeger  M., Mirimanoff  R. O.
ISSN
1013-2058 (Print)
Publication state
Published
Issued date
02/1991
Volume
80
Number
6
Pages
109-11
Notes
Case Reports
English Abstract
Journal Article --- Old month value: Feb 5
Abstract
A patient with cardiac pacing operated for cancer of the breast received subsequent radiotherapy, first with a linear accelerator, then with a betatron. The pacemaker was found deprogrammed as a sequel to irradiation although care was taken to keep it outside the irradiation field. After reprogramming the pacemaker changed a second time spontaneously the given parameters indicating permanent damage by the irradiation. In such cases the exchange of the pacemaker is recommended. Precautionary measures ranging from surveillance to possibilities for reprogramming and even reanimation or pacemaker replacement should be assured prior to irradiation of patients with cardiac pacing.
Keywords
Breast Neoplasms/*radiotherapy Equipment Failure Female Humans Middle Aged Monitoring, Physiologic *Pacemaker, Artificial Radiotherapy, High-Energy/*adverse effects