Accelerated MP2RAGE imaging using Cartesian phyllotaxis readout and compressed sensing reconstruction.

Details

Serval ID
serval:BIB_6D2E34072375
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Accelerated MP2RAGE imaging using Cartesian phyllotaxis readout and compressed sensing reconstruction.
Journal
Magnetic resonance in medicine
Author(s)
Mussard E., Hilbert T., Forman C., Meuli R., Thiran J.P., Kober T.
ISSN
1522-2594 (Electronic)
ISSN-L
0740-3194
Publication state
Published
Issued date
10/2020
Peer-reviewed
Oui
Volume
84
Number
4
Pages
1881-1894
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Abstract
MP2RAGE T <sub>1</sub> -weighted imaging has been shown to be beneficial for various applications, mainly because of its good grey-white matter contrast, its B <sub>1</sub> -robustness and ability to derive T <sub>1</sub> maps. Even using parallel imaging, the method requires long acquisition times, especially at high resolution. This work aims at accelerating MP2RAGE imaging using compressed sensing.
A pseudo-phyllotactic Cartesian MP2RAGE readout was implemented allowing for flexible reordering and undersampling factors. The sampling pattern was first optimized based on fully sampled data and a compressed sensing reconstruction. Changes in contrast ratios, automated brain segmentation results, and quantitative T <sub>1</sub> values were used for benchmarking. In vivo undersampled data from eleven healthy subjects were then acquired using a 4-fold acceleration with the optimized sampling pattern. The resulting images were compared to the standard parallel imaging MP2RAGE protocol by visual inspection and using the above quality metrics.
The application of incoherent undersampling and iterative compressed sensing reconstruction on MP2RAGE acquisitions allows for a 57% time reduction (corresponding to 4-fold undersampling with maintained reference lines, TA = 3:35 minutes) compared to the reference protocol using parallel imaging (GRAPPAx3 acceleration, TA = 8:22 minutes) while obtaining images with similar image quality, morphometric (volume differences = [0.07 ± 1.2-3.8 ± 1.9]%) and T <sub>1</sub> -mapping outcomes (T <sub>1</sub> error = [6 ± 5.1-37 ± 12.3] ms depending on the different structures).
A whole-brain MP2RAGE acquisition is feasible with compressed sensing in less than 4 minutes without appreciably compromising image quality.
Keywords
Brain/diagnostic imaging, Healthy Volunteers, Humans, Image Enhancement, Image Interpretation, Computer-Assisted, Magnetic Resonance Imaging, MP2RAGE, T1 map, compressed sensing, phyllotaxis pattern
Pubmed
Web of science
Create date
02/04/2020 17:06
Last modification date
20/07/2023 6:57