Assessment of UVA-Riboflavin Corneal Cross-Linking Using Small Amplitude Oscillatory Shear Measurements.

Details

Serval ID
serval:BIB_6D1AC721561E
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Assessment of UVA-Riboflavin Corneal Cross-Linking Using Small Amplitude Oscillatory Shear Measurements.
Journal
Investigative ophthalmology & visual science
Author(s)
Aslanides I.M., Dessi C., Georgoudis P., Charalambidis G., Vlassopoulos D., Coutsolelos A.G., Kymionis G., Mukherjee A., Kitsopoulos T.N.
ISSN
1552-5783 (Electronic)
ISSN-L
0146-0404
Publication state
Published
Issued date
01/04/2016
Peer-reviewed
Oui
Volume
57
Number
4
Pages
2240-2245
Language
english
Notes
Publication types: Journal Article ; Research Support, Non-U.S. Gov't
Publication Status: ppublish
Abstract
The effect of ultraviolet (UV)-riboflavin cross-linking (CXL) has been measured primarily using the strip extensometry technique. We propose a simple and reliable methodology for the assessment of CXL treatment by using an established rheologic protocol based on small amplitude oscillatory shear (SAOS) measurements. It provides information on the average cross-link density and the elastic modulus of treated cornea samples.
Three fresh postmortem porcine corneas were used to study the feasibility of the technique, one serving as control and two receiving corneal collagen cross-linking treatment. Subsequently, five pairs of fresh postmortem porcine corneas received corneal collagen cross-linking treatment with riboflavin and UVA-irradiation (370 nm; irradiance of 3 mW/cm2) for 30 minutes (Dresden protocol); the contralateral porcine corneas were used as control samples. After the treatment, the linear viscoelastic moduli of the corneal samples were measured using SAOS measurements and the average cross-linking densities extracted.
For all cases investigated, the dynamic moduli of the cross-linked corneas were higher compared to those of the corresponding control samples. The increase of the elastic modulus of the treated samples was between 122% and 1750%. The difference was statistically significant for all tested samples (P = 0.018, 2-tailed t-test).
We report a simple and accurate methodology for quantifying the effects of cross-linking on porcine corneas treated with the Dresden protocol by means of SAOS measurements in the linear regime. The measured dynamic moduli, elastic and viscous modulus, represent the energy storage and energy dissipation, respectively. Hence, they provide a means to assess the changing physical properties of the cross-linked collagen networks after CXL treatment.
Keywords
Animals, Collagen/metabolism, Cornea/metabolism, Cornea/physiology, Cornea/radiation effects, Elasticity/physiology, Elasticity/radiation effects, Riboflavin/metabolism, Shear Strength/physiology, Shear Strength/radiation effects, Swine, Ultraviolet Rays, Viscosity/radiation effects