European Survey of Image Quality Assessment Methods Used in Mammography
Journal
Radiation Protection Dosimetry
Author(s)
Verdun F.R., Moeckli R., Valley J.F., Bochud F., Hessler C., Schnyder P.
ISSN
0144-8420
Publication state
Published
Issued date
1998
Peer-reviewed
Oui
Volume
80
Number
1-3
Pages
73-76
Language
english
Notes
Document Type: Article; Proceedings Paper
Reference doses and quality in medical imaging:what the referring practitioner and directing medical staff should know : proceedings of a workshop, Luxembourg, October 23-25, 1997
Abstract
The definition of reference dose levels has to be linked with the definition of image quality. Unfortunately, there is
still no general agreement on the definition of image quality in mammography, and most of the protocols used are based on the
detectability of objects having various shapes and contrasts. To facilitate the task of assessing image quality, scoring methods
are often used to produce a single number representative of the imaging chain performance. The goal of this study is to present
a comparison between different ways of assessing image quality commonly used in Europe. A set of five mammograms, having
different image quality levels, has been obtained with several test objects and compared. The results show large sensitivity
variations among the different methods. Concerted work between radiologists and physicists is still required to define the radiological
tasks and develop objective ways to measure image quality in mammography.