Validation of virtual fractional flow reserve pullback curves.

Details

Serval ID
serval:BIB_57BC395921FC
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Validation of virtual fractional flow reserve pullback curves.
Journal
Catheterization and cardiovascular interventions
Author(s)
Seki R., Collison D., Ikeda K., Sonck J., Munhoz D., Bertolone D.T., Ko B., Maeng M., Otake H., Koo B.K., Storozhenko T., Bouisset F., Belmonte M., Leone A., Shumkova M., Ford T.J., Mahendiran T., Berry C., De Bruyne B., Oldroyd K., Sakai K., Mizukami T., Collet C.
ISSN
1522-726X (Electronic)
ISSN-L
1522-1946
Publication state
Published
Issued date
11/2024
Peer-reviewed
Oui
Volume
104
Number
6
Pages
1178-1188
Language
english
Notes
Publication types: Journal Article ; Validation Study ; Comparative Study
Publication Status: ppublish
Abstract
Angiography-derived fractional flow reserve (virtual FFR) has shown excellent diagnostic performance compared with wire-based FFR. However, virtual FFR pullback curves have not been validated yet.
To validate the accuracy of virtual FFR pullback curves compared to wire-based FFR pullbacks and to assess their clinical utility using patient-reported outcomes.
Pooled analysis of two prospective studies, including patients with hemodynamically significant (FFR ≤ 0.80) coronary artery disease (CAD). Virtual and wire-based FFR pullbacks were compared to assess the accuracy of virtual pullbacks to characterize CAD as focal or diffuse. Pullbacks were analyzed visually and quantitatively using the pullback pressure gradient (PPG). Patients underwent PCI, and the Seattle Angina Questionnaire (SAQ) was administered at 3-month follow-up.
A total of 298 patients (300 vessels) with both virtual and wire-based pullbacks who underwent PCI were included in the analysis. The mean age was 61.8 ± 8.8, and 15% were female. The agreement on the visual adjudication of the CAD pattern was fair (Cohen's Kappa: 0.31, 95% confidence interval: 0.18-0.45). The mean PPG were 0.65 ± 0.18 from virtual pullbacks and 0.65 ± 0.13 from wire-based pullbacks (r = 0.68, mean difference 0, limits of agreement -0.27 to 0.28). At follow-up, patients with high virtual PPG (>0.67) had higher SAQ angina frequency scores (i.e., less angina) than those with low virtual PPG (SAQ scores 92.0 ± 14.3 vs. 85.5 ± 23.1, p = 0.022).
Virtual FFR pullback curves showed moderate agreement with wire-based FFR pullbacks. Nonetheless, patients with focal disease based on virtual PPG reported greater improvement in angina after PCI.
Keywords
Humans, Fractional Flow Reserve, Myocardial, Female, Male, Middle Aged, Coronary Artery Disease/physiopathology, Coronary Artery Disease/diagnostic imaging, Coronary Artery Disease/therapy, Predictive Value of Tests, Aged, Reproducibility of Results, Coronary Angiography, Percutaneous Coronary Intervention/adverse effects, Cardiac Catheterization, Treatment Outcome, Coronary Vessels/physiopathology, Coronary Vessels/diagnostic imaging, Time Factors, Prospective Studies, Angina Pectoris/physiopathology, Angina Pectoris/therapy, Angina Pectoris/diagnostic imaging, Angina Pectoris/etiology, Severity of Illness Index, Surveys and Questionnaires, PCI, angina, angiography‐derived fractional flow reserve, coronary artery disease, fractional flow reserve
Pubmed
Web of science
Create date
04/10/2024 15:38
Last modification date
20/11/2024 7:16
Usage data