Validation of a tonometric noninvasive arterial blood pressure monitor in the intensive care setting.

Details

Serval ID
serval:BIB_54C3CAE4231B
Type
Article: article from journal or magazin.
Collection
Publications
Title
Validation of a tonometric noninvasive arterial blood pressure monitor in the intensive care setting.
Journal
Anaesthesia
Author(s)
Steiner L.A., Johnston A.J., Salvador R., Czosnyka M., Menon D.K.
ISSN
0003-2409
Publication state
Published
Issued date
2003
Peer-reviewed
Oui
Volume
58
Number
5
Pages
448-454
Language
english
Notes
Publication types: Comparative Study ; Journal Article ; Research Support, Non-U.S. Gov't ; Validation Studies
Publication Status: ppublish
Abstract
Intra-arterial measurement is considered the gold standard for continuous, beat-to-beat arterial blood pressure monitoring. However, arterial cannulation can be difficult and may cause complications such as thrombosis and ischaemia. Recently, a tonometric system, the Colin CBM-7000 has been developed for noninvasive beat-to-beat measurement of arterial blood pressure from the radial artery. We assessed the level of agreement between the CBM-7000 and invasive radial artery measurements in 15 patients on a neuro-intensive care unit. Agreement of systolic, diastolic and mean arterial pressure values was limited, with approximately 34% of mean arterial pressures differing by over 10 mmHg. In many cases, this was due to a downward drift of the noninvasive measurements over time. Furthermore, there was a tendency to underestimate low pressures and overestimate high pressures. In our opinion, the Colin CBM-7000 cannot be recommended for continuous blood pressure monitoring in the intensive care setting.
Keywords
Adult, Blood Pressure Monitors, Humans, Intensive Care/methods, Manometry/instrumentation, Monitoring, Physiologic/instrumentation, Reproducibility of Results