Transvenous Lead Extraction during Cardiac Implantable Device Upgrade: Results from the Multicenter Swiss Lead Extraction Registry.

Details

Ressource 1Download: 37629216_BIB_4959BB265EB0.pdf (1451.79 [Ko])
State: Public
Version: Final published version
License: CC BY 4.0
Serval ID
serval:BIB_4959BB265EB0
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Transvenous Lead Extraction during Cardiac Implantable Device Upgrade: Results from the Multicenter Swiss Lead Extraction Registry.
Journal
Journal of clinical medicine
Author(s)
Haeberlin A., Noti F., Breitenstein A., Auricchio A., Reichlin T., Conte G., Klersy C., Curti M., Pruvot E., Domenichini G., Schaer B., Kühne M., Gruszczynski M., Burri H., Kobza R., Grebmer C., Regoli F.D.
ISSN
2077-0383 (Print)
ISSN-L
2077-0383
Publication state
Published
Issued date
31/08/2023
Peer-reviewed
Oui
Volume
12
Number
16
Pages
5175
Language
english
Notes
Publication types: Journal Article
Publication Status: epublish
Abstract
Device patients may require upgrade interventions from simpler to more complex cardiac implantable electronic devices. Prior to upgrading interventions, clinicians need to balance the risks and benefits of transvenous lead extraction (TLE), additional lead implantation or lead abandonment. However, evidence on procedural outcomes of TLE at the time of device upgrade is scarce.
This is a post hoc analysis of the investigator-initiated multicenter Swiss TLE registry. The objectives were to assess patient and procedural factors influencing TLE outcomes at the time of device upgrades.
941 patients were included, whereof 83 (8.8%) had TLE due to a device upgrade. Rotational mechanical sheaths were more often used in upgraded patients (59% vs. 42.7%, p = 0.015) and total median procedure time was longer in these patients (160 min vs. 105 min, p < 0.001). Clinical success rates of upgraded patients compared to those who received TLE due to other reasons were not different (97.6% vs. 93.0%, p = 0.569). Moreover, multivariable analysis showed that upgrade procedures were not associated with a greater risk for complications (HR 0.48, 95% confidence interval 0.14-1.57, p = 0.224; intraprocedural complication rate of upgraded patients 7.2% vs. 5.5%). Intraprocedural complications of upgraded patients were mostly associated with the implantation and not the extraction procedure (67% vs. 33% of complications).
TLE during device upgrade is effective and does not attribute a disproportionate risk to the upgrade procedure.
Keywords
lead extraction complications, lead extraction indication, lead extraction risk factors, transvenous lead extraction, upgrade procedure