Optimization of Alanine Measurements for Fast and Accurate Dosimetry in FLASH Radiation Therapy.

Details

Ressource 1Request a copy Under indefinite embargo.
UNIL restricted access
State: Public
Version: author
License: Not specified
Serval ID
serval:BIB_47E03865D8D3
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Optimization of Alanine Measurements for Fast and Accurate Dosimetry in FLASH Radiation Therapy.
Journal
Radiation research
Author(s)
Gondré M., Jorge P.G., Vozenin M.C., Bourhis J., Bochud F., Bailat C., Moeckli R.
ISSN
1938-5404 (Electronic)
ISSN-L
0033-7587
Publication state
Published
Issued date
01/12/2020
Peer-reviewed
Oui
Volume
194
Number
6
Pages
573-579
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Abstract
FLASH radiation therapy (FLASH-RT) reference dosimetry to obtain traceability, repeatability and stability of irradiations cannot be performed with conventional dosimetric methods, such as monitor chambers or ionization chambers. Until now, only passive dosimeters have provided the necessary dosimetric data. Alanine dosimetry is accurate; however, to be used for FLASH-RT in biological experiments and for clinical transfer to humans, the reading time needs to be reduced, while preserving a maximum deviation to the reference of ±2%. Optimization of alanine dosimetry was based on the acquisition of electron paramagnetic resonance (EPR) spectra with a Bruker spectrometer. Reading parameters such as the conversion time, the number of scans, the time constant, the microwave power and the modulation amplitude of the magnetic field were optimized as a trade-off between the signal-to-noise ratio (SNR) and the reading time of one measurement using the reference 10.1 Gy alanine pellet. After optimizing the parameters, we compared the doses measured with alanine pellets up to 100 Gy with the reference doses, and then determined the number of measurements necessary to get a difference lower than ±2%. A low-dose alanine pellet of 4.9 Gy was also measured to evaluate the quality of the optimization for doses lower than 10 Gy. The optimization of the Bruker default parameters made it possible to reduce the reading time for one measurement from 5.6 to 2.6 min. That reduction was not at the cost of the SNR because it was kept comparable to the default parameters. Three measurements were enough to obtain a maximum dose deviation to the reference of 1.8% for the range of 10-100 Gy. The total reading time for the three measurements was 7.8 min (3 × 2.6 min). For lower doses such as 4.9 Gy, three measurements led to a deviation greater than 5%. By increasing the number of measurements to five, the average difference to the reference dose was reduced to less than 5% with a total reading time increased to 13.0 min. For doses between 10 Gy and 100 Gy, the optimized acquisition parameters made it possible to keep the average differences between the reference and the measured doses below ±2%, for a reading time of 7.8 min. This enabled an accurate and fast dose determination for biological preparations as part of FLASH-beam irradiations.
Keywords
Alanine/analysis, Electron Spin Resonance Spectroscopy/methods, Humans, Radiotherapy/methods, Radiotherapy Dosage
Pubmed
Web of science
Open Access
Yes
Create date
10/06/2020 22:03
Last modification date
14/06/2024 6:02
Usage data