Reproducibility of straylight measurement by C-Quant for assessment of retinal straylight using the compensation comparison method.

Details

Ressource 1Download: REF.pdf (146.48 [Ko])
State: Public
Version: Final published version
License: Not specified
It was possible to publish this article open access thanks to a Swiss National Licence with the publisher.
Serval ID
serval:BIB_438A346F510B
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Reproducibility of straylight measurement by C-Quant for assessment of retinal straylight using the compensation comparison method.
Journal
Graefe's Archive For Clinical and Experimental Ophthalmology = Albrecht Von Graefes Archiv Für Klinische Und Experimentelle Ophthalmologie
Author(s)
Guber I., Bachmann L.M., Guber J., Bochmann F., Lange A.P., Thiel M.A.
ISSN
1435-702X (Electronic)
ISSN-L
0721-832X
Publication state
Published
Issued date
2011
Peer-reviewed
Oui
Volume
249
Number
9
Pages
1367-1371
Language
english
Notes
Publication types: Journal ArticlePublication Status: ppublish
Abstract
BACKGROUND: Straylight gives the appearance of a veil of light thrown over a person's retinal image when there is a strong light source present. We examined the reproducibility of the measurements by C-Quant, and assessed its correlation to characteristics of the eye and subjects' age. PARTICIPANTS AND METHODS: Five repeated straylight measurements were taken using the dominant eye of 45 healthy subjects (age 21-59) with a BCVA of 20/20: 14 emmetropic, 16 myopic, eight hyperopic and seven with astigmatism. We assessed the extent of reproducibility of straylight measures using the intraclass correlation coefficient. RESULTS: The mean straylight value of all measurements was 1.01 (SD 0.23, median 0.97, interquartile range 0.85-1.1). Per 10 years of age, straylight increased in average by 0.10 (95%CI 0.04 to 0.16, p < 0.01]. We found no independent association of refraction (range -5.25 dpt to +2 dpt) on straylight values (0.001; 95%CI -0.022 to 0.024, p = 0.92). Compared to emmetropic subjects, myopia reduced straylight (-.011; -0.024 to 0.02, p = 0.11), whereas higher straylight values (0.09; -0.01 to 0.20, p = 0.09) were observed in subjects with blue irises as compared to dark-colored irises when correcting for age. The intraclass correlation coefficient (ICC) of repeated measurements was 0.83 (95%CI 0.76 to 0.90). CONCLUSIONS: Our study showed that straylight measurements with the C-Quant had a high reproducibility, i.e. a lack of large intra-observer variability, making it appropriate to be applied in long-term follow-up studies assessing the long-term effect of surgical procedures on the quality of vision.