Development and experimental validation of a finite element model of total ankle replacement.

Details

Serval ID
serval:BIB_377C5B4EC574
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Development and experimental validation of a finite element model of total ankle replacement.
Journal
Journal of Biomechanics
Author(s)
Terrier A., Larrea X., Guerdat J., Crevoisier X.
ISSN
1873-2380 (Electronic)
ISSN-L
0021-9290
Publication state
Published
Issued date
2014
Peer-reviewed
Oui
Volume
47
Number
3
Pages
742-745
Language
english
Notes
Publication types: Journal Article ; pdf : Short communication
Abstract
Total ankle replacement remains a less satisfactory solution compared to other joint replacements. The goal of this study was to develop and validate a finite element model of total ankle replacement, for future testing of hypotheses related to clinical issues. To validate the finite element model, an experimental setup was specifically developed and applied on 8 cadaveric tibias. A non-cemented press fit tibial component of a mobile bearing prosthesis was inserted into the tibias. Two extreme anterior and posterior positions of the mobile bearing insert were considered, as well as a centered one. An axial force of 2kN was applied for each insert position. Strains were measured on the bone surface using digital image correlation. Tibias were CT scanned before implantation, after implantation, and after mechanical tests and removal of the prosthesis. The finite element model replicated the experimental setup. The first CT was used to build the geometry and evaluate the mechanical properties of the tibias. The second CT was used to set the implant position. The third CT was used to assess the bone-implant interface conditions. The coefficient of determination (R-squared) between the measured and predicted strains was 0.91. Predicted bone strains were maximal around the implant keel, especially at the anterior and posterior ends. The finite element model presented here is validated for future tests using more physiological loading conditions.
Pubmed
Web of science
Create date
21/04/2014 16:49
Last modification date
20/08/2019 14:25
Usage data