A new statistical methodology overcame the defects of the Bland-Altman method.

Details

Ressource 1Download: A new statistical methodology overcame the defects of the Bland and Altman method (version accepté).pdf (1750.38 [Ko])
State: Public
Version: Final published version
License: Not specified
Serval ID
serval:BIB_36740603C2D3
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
A new statistical methodology overcame the defects of the Bland-Altman method.
Journal
Journal of clinical epidemiology
Author(s)
Taffé P., Halfon P., Halfon M.
ISSN
1878-5921 (Electronic)
ISSN-L
0895-4356
Publication state
Published
Issued date
08/2020
Peer-reviewed
Oui
Volume
124
Pages
1-7
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Abstract
The Bland and Altman's limits of agreement (LoA) method is the most commonly used statistical method to assess the bias and precision of a new measuring device (it has been cited over 40,000 times as of March 2019). What is less known is that the LoA method can be dramatically misleading.
A new statistical methodology, which circumvents these deficiencies, has recently been published and made available in the R and Stata statistical packages. We aimed at introducing and illustrating with a small data set on blood pressure (BP) measurements, taken by two different oscillometric devices, the use of this new methodology to a clinical audience.
For diastolic BP, the LoA method was particularly misleading, as it identified differential and proportional biases of opposite signs compared with the new methodology. Regarding systolic BP, the LoA method strongly overestimated both the differential and proportional biases, for both devices.
The LoA method may be dramatically misleading and does not allow one to estimate the precision of each measurement method. We recommend the use of the newly developed statistical methodology instead.
Keywords
Biometry/methods, Blood Pressure, Blood Pressure Determination, Humans, Oscillometry, Research Design, Bias, Blood pressure, Limits of Agreement, Oscillometric device, Precision
Pubmed