Construction and dosimetric characterization of a motorized scanning-slit system for electron FLASH experiments.

Details

Ressource 1Download: 37439505.pdf (992.50 [Ko])
State: Public
Version: Final published version
License: CC BY 4.0
Serval ID
serval:BIB_30DA34647F67
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Construction and dosimetric characterization of a motorized scanning-slit system for electron FLASH experiments.
Journal
Medical physics
Author(s)
Oesterle R., Bailat C., Buhlmann D., Bochud F., Grilj V.
ISSN
2473-4209 (Electronic)
ISSN-L
0094-2405
Publication state
Published
Issued date
02/2024
Peer-reviewed
Oui
Volume
51
Number
2
Pages
1396-1404
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Abstract
Beam scanning is a useful technique for the treatment of large tumors when the primary beam size is limited, which is the case with radiation beams used in FLASH radiotherapy.
To optimize beam scanning as a dose delivery method for FLASH radiotherapy, it is necessary to first understand the effects of beam scanning on the FLASH effect. To do so, biological FLASH experiments need to be done using defined beam parameters with beam scanning and compared to the situation without beam scanning. In this regard, we propose implementation of a simple slit scanning system with an electron FLASH beam to obtain a scanned radiation field that closely resembles a static field.
A pulsed electron linear accelerator (linac) was used in combination with a scanning slit system in order to simulate a scanned electron beam. Three configurations that produced homogeneous lateral profiles and high enough doses per pulse for FLASH experiments were established. The optimal scanning parameters were found for each configuration by examining the flatness of the obtained lateral dose profiles. Using the optimal scanning parameters, the scanned FLASH beams were dosimetrically characterized and compared to non-scanned open field beam.
A final electron FLASH beam scanning configuration was found for a 1 mm wide slit at a distance of 350 mm from the linac and a 2 mm wide slit at distances of 350 and 490 mm from the linac. The lateral profiles for these final configurations were found to have a homogeneity that is comparable to the open field profiles. The percentage depth dose (PDD) values found for these final configurations closely matched (by a few percentage) the PDD of the open field beam.
Three electron FLASH beam scanning configurations achieved by the motorized slit system were found to produce radiation fields similar to a non-scanned open field electron beam. These final configurations can therefore be used in future biological FLASH experiments to compare to non-scanned beam experiments in order to optimize beam scanning as a technique permitting the treatment of larger tumors with FLASH radiotherapy.
Keywords
FLASH, beam scanning, radiotherapy, slit collimator
Pubmed
Web of science
Open Access
Yes
Create date
14/07/2023 13:41
Last modification date
13/02/2024 8:28
Usage data