Evaluation of organ-specific peripheral doses after 2-dimensional, 3-dimensional and hybrid intensity modulated radiation therapy for breast cancer based on Monte Carlo and convolution/superposition algorithms: implications for secondary cancer risk assessment.

Details

Serval ID
serval:BIB_2E35C2D6242A
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Evaluation of organ-specific peripheral doses after 2-dimensional, 3-dimensional and hybrid intensity modulated radiation therapy for breast cancer based on Monte Carlo and convolution/superposition algorithms: implications for secondary cancer risk assessment.
Journal
Radiotherapy and oncology
Author(s)
Joosten A., Matzinger O., Jeanneret-Sozzi W., Bochud F., Moeckli R.
ISSN
1879-0887 (Electronic)
ISSN-L
0167-8140
Publication state
Published
Issued date
01/2013
Peer-reviewed
Oui
Volume
106
Number
1
Pages
33-41
Language
english
Notes
Publication types: Journal Article ; Research Support, Non-U.S. Gov't
Publication Status: ppublish
Abstract
To make a comprehensive evaluation of organ-specific out-of-field doses using Monte Carlo (MC) simulations for different breast cancer irradiation techniques and to compare results with a commercial treatment planning system (TPS).
Three breast radiotherapy techniques using 6MV tangential photon beams were compared: (a) 2DRT (open rectangular fields), (b) 3DCRT (conformal wedged fields), and (c) hybrid IMRT (open conformal+modulated fields). Over 35 organs were contoured in a whole-body CT scan and organ-specific dose distributions were determined with MC and the TPS.
Large differences in out-of-field doses were observed between MC and TPS calculations, even for organs close to the target volume such as the heart, the lungs and the contralateral breast (up to 70% difference). MC simulations showed that a large fraction of the out-of-field dose comes from the out-of-field head scatter fluence (>40%) which is not adequately modeled by the TPS. Based on MC simulations, the 3DCRT technique using external wedges yielded significantly higher doses (up to a factor 4-5 in the pelvis) than the 2DRT and the hybrid IMRT techniques which yielded similar out-of-field doses.
In sharp contrast to popular belief, the IMRT technique investigated here does not increase the out-of-field dose compared to conventional techniques and may offer the most optimal plan. The 3DCRT technique with external wedges yields the largest out-of-field doses. For accurate out-of-field dose assessment, a commercial TPS should not be used, even for organs near the target volume (contralateral breast, lungs, heart).

Keywords
Algorithms, Breast Neoplasms/radiotherapy, Female, Humans, Monte Carlo Method, Neoplasms, Radiation-Induced/etiology, Neoplasms, Second Primary/etiology, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated/adverse effects, Risk Assessment