Urea removal by neonatal hemo(dia)filters in hyperazotemic rabbits.

Details

Serval ID
serval:BIB_2D640CDD774F
Type
Article: article from journal or magazin.
Collection
Publications
Title
Urea removal by neonatal hemo(dia)filters in hyperazotemic rabbits.
Journal
Intensive Care Medicine
Author(s)
Gouyon J.B., Françoise M., Semama D.S., Prévot A., d'Athis P., Lallemant C.
ISSN
0342-4642
Publication state
Published
Issued date
1997
Peer-reviewed
Oui
Volume
23
Number
6
Pages
677-683
Language
english
Abstract
OBJECTIVE: Comparison of three neonatal hemo(dia)filters (FH22, Gambro; Minifilter Plus, Amicon; Miniflow 10, Hospal) for removal of urea by venovenous hemofiltration (HF) and venovenous hemodiafiltration (HDF). DESIGN: Filters were successively used for HF with two different blood flows (Qb = 8.3 and 16.6 ml/ min) and for HDF with the two different blood flows and four dialysate flows (Qd = 0.5, 1.0, 2.0, and 3.0 l/h). SUBJECTS: 21 anesthetized adult New Zealand White rabbits infused with urea. MAIN RESULTS: Urea clearance was significantly increased by HDF compared to HF regardless of blood flow, dialysate flow, and the hemo (dia)filter type except in the FH22 group, when blood flow was high and dialysate flow was 0.5 or 1.0 l/h. The FH22 filter allowed the best urea clearance during HF at high blood flow. During the HDF procedures, the Miniflow 10 allowed the highest urea clearance regardless of blood flow and dialysate flow.
Keywords
Analysis of Variance, Animals, Hemodiafiltration/instrumentation, Hemofiltration/instrumentation, Linear Models, Male, Rabbits, Urea/blood
Pubmed
Create date
20/07/2009 14:37
Last modification date
20/08/2019 14:12
Usage data