Self-navigated coronary MR angiography for coronary aneurysm detection in Kawasaki disease at 3T: comparison with conventional diaphragm-navigated coronary MR angiography.

Details

Serval ID
serval:BIB_29341D3FC3C6
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Self-navigated coronary MR angiography for coronary aneurysm detection in Kawasaki disease at 3T: comparison with conventional diaphragm-navigated coronary MR angiography.
Journal
European radiology
Author(s)
Zhou Z., Wei D., Azhe S., Fu C., Zhou X., An J., Piccini D., Bastiaansen J., Guo Y., Wen L.
ISSN
1432-1084 (Electronic)
ISSN-L
0938-7994
Publication state
In Press
Peer-reviewed
Oui
Language
english
Notes
Publication types: Journal Article
Publication Status: aheadofprint
Abstract
To assess the scan time, image quality, and diagnostic performance of self-navigated coronary MR angiography (SN-CMRA) for coronary aneurysm (CAA) detection in Kawasaki disease (KD) patients and compare it with diaphragm-navigated CMRA (DN-CMRA).
SN-CMRA and DN-CMRA were performed on 76 pediatric patients with KD (48 males, 6.75 ± 3.59 years). Thirty-three of whom underwent coronary CT angiography (CCTA)/invasive coronary angiography (ICA). The scan time and qualitative and quantitative image quality assessment were compared between the two sequences. The diagnostic performance for CAA detection by the two approaches using CCTA/ICA as the reference standard was compared on per-patient, per-vessel, and per-segment basis.
The scan time of SN-CMRA was significantly shorter than that of DN-CMRA (7.49 ± 2.31 min vs. 10.03 ± 4.47 min, p < 0.001). There was no difference in overall and segmental image quality to reach the clinical diagnostic criteria between the two sequences (all p > 0.05). No significant difference in vessel length of the three main coronary arteries was found between the two approaches (all p > 0.05). Moreover, SN-CMRA showed no difference from DN-CMRA in contrast ratio of blood-myocardium (1.25 (interquartile range [IQR], 1.06 to 1.51) vs. 1.18 (IQR, 0.95 to 1.64), p = 0.706). There was no difference in the diagnostic accuracy of SN-CMRA and DN-CMRA for CAA detection on per-patient, per-vessel, or per-segment basis (all p > 0.05).
SN-CMRA at 3T showed reliable diagnostic performance and application value for CAA detection in children with KD. Compared with DN-CMRA, SN-CMRA can simplify the scanning procedure and shorten the scan time, achieving comparable image quality and diagnostic accuracy.
Coronary aneurysm in children with Kawasaki disease (KD) can be detected by self-navigated coronary MR angiography (CMRA) non-invasively and without radiation, achieving comparable image quality and diagnostic performance as diaphragm-navigated CMRA while shortening scanning time. It can provide reference for risk stratification and treatment management of KD.
• Evaluating the size of coronary aneurysm is important for risk stratification and treatment of Kawasaki disease. • Self-navigated coronary MR angiography (SN-CMRA) shortens scan time and achieves comparable image quality and diagnostic performance compared with diaphragm-navigated coronary MR angiography. • SN-CMRA can evaluate coronary aneurysm non-invasively and without radiation, providing information for risk stratification and treatment.
Keywords
Coronary aneurysm, Coronary angiography, Magnetic resonance angiography, Mucocutaneous lymph node syndrome
Pubmed
Web of science
Create date
23/10/2023 12:28
Last modification date
14/11/2023 8:09
Usage data