Measurement of CBF and CMRO2 using the continuous inhalation of C15O2 and 15O. Experimental validation using CO2 reactivity in the anaesthetised dog.

Details

Serval ID
serval:BIB_24B5C30F2BBC
Type
Article: article from journal or magazin.
Collection
Publications
Title
Measurement of CBF and CMRO2 using the continuous inhalation of C15O2 and 15O. Experimental validation using CO2 reactivity in the anaesthetised dog.
Journal
Journal of the Neurological Sciences
Author(s)
Rhodes C.G., Lenzi G.L., Frackowiak R.S., Jones T., Pozzilli C.
ISSN
0022-510X (Print)
ISSN-L
0022-510X
Publication state
Published
Issued date
1981
Volume
50
Number
3
Pages
381-389
Language
english
Notes
Publication types: Comparative Study ; Journal ArticlePublication Status: ppublish
Abstract
Experimental support for the steady state 15O inhalation technique, as used to measure cerebral blood flow (CBF) and oxygen utilisation (CMRO2), was obtained by describing the response of the cerebral vasculature to variations in arterial PCO2 in 6 anaesthetised dogs. Measurements were made using a positron emission tomography (ECAT II) and arterial blood sampling, during the sequential constant inhalation of C15O2 and 15O2. Values of CBF and CMRO2 were calculated for a mixture of white and grey matter, using the steady state tracer equations derived by Jones et al. (1976). The mean CMRO2 was 3.58 +/- 0.81 ml O2 . 100 ml-1 . min-1, whilst the mean CBF and OER (oxygen extraction ratio) values (for an arterial PCO2 of 40 mm Hg) were 39.9 ml . 100 ml-1 and 0.50 +/- 0.06, respectively. Arterial PCO2 was varied between 20 and 150 mm Hg. CBF was found to correlate closely with arterial PCO2, resulting in a mean slope (specific reactivity) of 1.52 +/- 0.38 ml . 10 ml-1. mm Hg-1. Pooling the flow data resulted in a linear relationship between CBF (% change) and arterial PCO2 in the range 20-70 mm Hg, with a slope (% reactivity) of 3.2% mm Hg-1 (2 P less than 0.001). The oxygen extraction ratio (OER) fell with increasing values of arterial PCO2 resulting in a stable CMRO2 throughout each study. There was no correlation between CMRO2 and artificially increased CBF. These results support and give confidence in the use of the 15O inhalation technique for measuring CBF, OER and CMRO2.
Keywords
Animals, Carbon Dioxide/blood, Cerebrovascular Circulation, Dogs, Humans, Macaca mulatta, Oxygen, Oxygen Consumption, Partial Pressure, Physiology/methods, Regression Analysis, Tomography, Emission-Computed
Pubmed
Web of science
Create date
09/10/2011 19:42
Last modification date
20/08/2019 14:02
Usage data