A simple and rapid method for toxicological screening of 25 antidepressants in blood or urine using high performance liquid chromatography with diode-array detection.

Details

Serval ID
serval:BIB_1EBB05E5840A
Type
Article: article from journal or magazin.
Collection
Publications
Title
A simple and rapid method for toxicological screening of 25 antidepressants in blood or urine using high performance liquid chromatography with diode-array detection.
Journal
Annales De Biologie Clinique
Author(s)
Tracqui A., Kintz P., Kreissig P., Mangin P.
ISSN
0003-3898[print], 0003-3898[linking]
Publication state
Published
Issued date
1992
Volume
50
Number
9
Pages
639-647
Language
english
Abstract
A high performance liquid chromatographic method with diode-array detection (HPLC/DAD) for simultaneous screening of 25 antidepressants is presented. After single-step liquid-liquid extraction at pH9.5 using chloroform/2-propanol/n-heptane (60/14/26, v/v), the substances are eluted on a Novapak C18 4-microns column (300 x 3.9 mm, id), with methanol/tetrahydrofuran/pH 2.6 phosphate buffer (65/5/30, v/v) as the mobile phase (flow rate 0.8 ml/min). Full UV spectra from 200 to 400 nm are recorded on-line during the entire analysis and may be automatically compared to spectra stored in a library. The analysis is performed in 14 min. The method is simple, rapid, and highly specific.
Keywords
Antidepressive Agents/blood, Antidepressive Agents/classification, Chromatography, High Pressure Liquid/methods, Humans, Poisoning/blood, Poisoning/urine, Spectrophotometry, Ultraviolet
Pubmed
Create date
18/10/2010 13:58
Last modification date
20/08/2019 12:54
Usage data