Full-Body MRI in Patients With an Implantable Cardioverter-Defibrillator: Primary Results of a Randomized Study.

Details

Serval ID
serval:BIB_1CC703B81612
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Full-Body MRI in Patients With an Implantable Cardioverter-Defibrillator: Primary Results of a Randomized Study.
Journal
Journal of the American College of Cardiology
Author(s)
Gold M.R., Sommer T., Schwitter J., Al Fagih A., Albert T., Merkely B., Peterson M., Ciuffo A., Lee S., Landborg L., Cerkvenik J., Kanal E.
Working group(s)
Evera MRI Study Investigators
ISSN
1558-3597 (Electronic)
ISSN-L
0735-1097
Publication state
Published
Issued date
2015
Peer-reviewed
Oui
Volume
65
Number
24
Pages
2581-2588
Language
english
Abstract
BACKGROUND: Magnetic resonance imaging (MRI) of patients with conventional implantable cardioverter-defibrillators (ICD) is contraindicated.
OBJECTIVES: This multicenter, randomized trial evaluated safety and efficacy of a novel ICD system specially designed for full-body MRI without restrictions on heart rate or pacing dependency. The primary safety objective was >90% freedom from MRI-related events composite endpoint within 30 days post-MRI. The primary efficacy endpoints were ventricular pacing capture threshold and ventricular sensing amplitude.
METHODS: Subjects received either a single- or dual-chamber ICD. In a 2:1 randomization, subjects either underwent MRI at 1.5-T of the chest, cervical, and head regions to maximize radiofrequency exposure up to 2 W/kg specific absorption rate and gradient field exposure to 200 T/m/s per axis (MRI group, n = 175), or they underwent a 1-h waiting period without MRI (control group, n = 88). A subset of MRI patients underwent ventricular fibrillation induction testing post-MRI to characterize defibrillation function.
RESULTS: In 42 centers, 275 patients were enrolled (76% male, age 60.4 ± 13.8 years). The safety endpoint was met with 100% freedom from the composite endpoint (p < 0.0001). Both efficacy endpoints were met with minimal differences in the proportion of MRI and control patients who demonstrated a ≤0.5 V increase in ventricular pacing capture threshold (100% MRI vs. 98.8% control, noninferiority p < 0.0001) or a ≤50% decrease in R-wave amplitude (99.3% MRI vs. 98.8% control, noninferiority p = 0.0001). A total of 34 ventricular tachyarrhythmia/ventricular fibrillation episodes (20 induced; 14 spontaneous) occurred in 24 subjects post-MRI, with no observed effect on sensing, detection, or treatment.
CONCLUSIONS: This is the first randomized clinical study of an ICD system designed for full-body MRI at 1.5-T. These data support that the system is safe and the MRI scan does not adversely affect electrical performance or efficacy. (Confirmatory Clinical Trial of the Evera MRI System for Conditionally-Safe MRI Access; NCT02117414).
Pubmed
Web of science
Open Access
Yes
Create date
13/07/2015 11:20
Last modification date
20/08/2019 12:53