Calibration of surface contamination monitors for the detection of iodine incorporation in the thyroid gland.

Details

Ressource 1Download: REF.pdf (96.93 [Ko])
State: Public
Version: Final published version
License: Not specified
It was possible to publish this article open access thanks to a Swiss National Licence with the publisher.
Serval ID
serval:BIB_1B0F9E229D9B
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Calibration of surface contamination monitors for the detection of iodine incorporation in the thyroid gland.
Journal
Radiation Protection Dosimetry
Author(s)
Bailat C., Buchillier T., Baechler S., Bochud F.
ISSN
1742-3406 (Electronic)
ISSN-L
0144-8420
Publication state
Published
Issued date
2011
Volume
144
Number
1-4
Pages
505-509
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Abstract
In Switzerland, individuals exposed to the risk of activity intake are required to perform regular monitoring. Monitoring consists in a screening measurement and is meant to be performed using commonly available laboratory instruments. More particularly, iodine intake is measured using a surface contamination monitor. The goal of the present paper is to report the calibration method developed for thyroid screening instruments. It consists of measuring the instrument response to a known activity located in the thyroid gland of a standard neck phantom. One issue of this procedure remains that the iodine radioisotopes have a short half-life. Therefore, the adequacy and limitations to simulate the short-lived radionuclides with so-called mock radionuclides of longer half-life were also evaluated. In light of the results, it has been decided to use only the appropriate iodine sources to perform the calibration.
Keywords
Calibration, Equipment Contamination, Equipment Design, Humans, Iodine Radioisotopes/analysis, Models, Statistical, Occupational Exposure/prevention & control, Phantoms, Imaging, Polyethylene/chemistry, Radiation Monitoring/instrumentation, Radiation Monitoring/methods, Radiation Protection/instrumentation, Radiation Protection/methods, Reproducibility of Results, Sodium Iodide/chemistry, Switzerland, Thyroid Gland/radiation effects, Thyroid Gland/radionuclide imaging