Comparison of airway pressure-triggered and airflow-triggered ventilation in very immature infants

Details

Serval ID
serval:BIB_09A8A13EBC11
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Comparison of airway pressure-triggered and airflow-triggered ventilation in very immature infants
Journal
Acta Paediatrica
Author(s)
Dimitriou  G., Greenough  A., Laubscher  B., Yamaguchi  N.
ISSN
0803-5253 (Print)
Publication state
Published
Issued date
12/1998
Volume
87
Number
12
Pages
1256-60
Notes
Comparative Study
Journal Article
Research Support, Non-U.S. Gov't --- Old month value: Dec
Abstract
Failure of patient-triggered ventilation in very immature infants may be due to the use of inappropriate triggering systems. Two types of airflow trigger were therefore compared consecutively to an airway pressure (SLE) triggering system. Each comparison was made in 10 infants, < or =28 weeks of gestation. Comparison was made of the delivered volume, trigger performance and blood gases using each system for 1 h. Both comparisons showed that the airflow triggering systems performed better: one (Draeger Babylog 8000) had a higher sensitivity (p < 0.01) and the other (Bird VIP airflow trigger), in which inflation was terminated by sensing a reduction in inspiratory flow, had a lower degree of asynchrony (p < 0.01) and a tendency to deliver higher volumes. These results suggest that triggering systems sensing airflow changes may be superior to those sensing airway pressure changes in very immature infants. The use of a mechanism to synchronize the termination of inflation to the end of the patient's inspiration may offer further advantages.
Keywords
Humans Infant, Newborn *Infant, Premature Infant, Premature, Diseases/*therapy *Pulmonary Ventilation Respiration, Artificial/*methods Treatment Outcome