Report from the HarmoSter study: different LC-MS/MS androstenedione, DHEAS and testosterone methods compare well; however, unifying calibration is a double-edged sword.
Détails
Télécharger: 38205643 .pdf (747.55 [Ko])
Etat: Public
Version: Final published version
Licence: CC BY 4.0
Etat: Public
Version: Final published version
Licence: CC BY 4.0
ID Serval
serval:BIB_D83302727151
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Report from the HarmoSter study: different LC-MS/MS androstenedione, DHEAS and testosterone methods compare well; however, unifying calibration is a double-edged sword.
Périodique
Clinical chemistry and laboratory medicine
ISSN
1437-4331 (Electronic)
ISSN-L
1434-6621
Statut éditorial
Publié
Date de publication
27/05/2024
Peer-reviewed
Oui
Volume
62
Numéro
6
Pages
1080-1091
Langue
anglais
Notes
Publication types: Journal Article ; Comparative Study
Publication Status: epublish
Publication Status: epublish
Résumé
Current liquid chromatography-tandem mass spectrometry (LC-MS/MS) applications for circulating androgen measurements are technically diverse. Previously, variable results have been reported for testosterone. Data are scarce for androstenedione and absent for dehydroepiandrosterone sulfate (DHEAS). We assessed the agreement of androstenedione, DHEAS and testosterone LC-MS/MS measurements among nine European centers and explored benefits of calibration system unification.
Androgens were measured twice by laboratory-specific procedures in 78 patient samples and in EQA materials. Results were obtained by in-house and external calibration. Intra- and inter-laboratory performances were valued.
Intra-laboratory CVs ranged between 4.2-13.2 % for androstenedione, 1.6-10.8 % for DHEAS, and 4.3-8.7 % and 2.6-7.1 % for female and male testosterone, respectively. Bias and trueness in EQA materials were within ±20 %. Median inter-laboratory CV with in-house vs. external calibration were 12.0 vs. 9.6 % for androstenedione (p<0.001), 7.2 vs. 4.9 % for DHEAS (p<0.001), 6.4 vs. 7.6 % for female testosterone (p<0.001) and 6.8 and 7.4 % for male testosterone (p=0.111). Median bias vs. all laboratory median with in-house and external calibration were -13.3 to 20.5 % and -4.9 to 18.7 % for androstenedione, -10.9 to 4.8 % and -3.4 to 3.5 % for DHEAS, -2.7 to 6.5 % and -11.3 to 6.6 % for testosterone in females, and -7.0 to 8.5 % and -7.5 to 11.8 % for testosterone in males, respectively.
Methods showed high intra-laboratory precision but variable bias and trueness. Inter-laboratory agreement was remarkably good. Calibration system unification improved agreement in androstenedione and DHEAS, but not in testosterone measurements. Multiple components, such as commutability of calibrators and EQA materials and internal standard choices, likely contribute to inter-laboratory variability.
Androgens were measured twice by laboratory-specific procedures in 78 patient samples and in EQA materials. Results were obtained by in-house and external calibration. Intra- and inter-laboratory performances were valued.
Intra-laboratory CVs ranged between 4.2-13.2 % for androstenedione, 1.6-10.8 % for DHEAS, and 4.3-8.7 % and 2.6-7.1 % for female and male testosterone, respectively. Bias and trueness in EQA materials were within ±20 %. Median inter-laboratory CV with in-house vs. external calibration were 12.0 vs. 9.6 % for androstenedione (p<0.001), 7.2 vs. 4.9 % for DHEAS (p<0.001), 6.4 vs. 7.6 % for female testosterone (p<0.001) and 6.8 and 7.4 % for male testosterone (p=0.111). Median bias vs. all laboratory median with in-house and external calibration were -13.3 to 20.5 % and -4.9 to 18.7 % for androstenedione, -10.9 to 4.8 % and -3.4 to 3.5 % for DHEAS, -2.7 to 6.5 % and -11.3 to 6.6 % for testosterone in females, and -7.0 to 8.5 % and -7.5 to 11.8 % for testosterone in males, respectively.
Methods showed high intra-laboratory precision but variable bias and trueness. Inter-laboratory agreement was remarkably good. Calibration system unification improved agreement in androstenedione and DHEAS, but not in testosterone measurements. Multiple components, such as commutability of calibrators and EQA materials and internal standard choices, likely contribute to inter-laboratory variability.
Mots-clé
Androstenedione/blood, Androstenedione/analysis, Testosterone/blood, Testosterone/analysis, Testosterone/standards, Humans, Tandem Mass Spectrometry/standards, Tandem Mass Spectrometry/methods, Calibration, Male, Female, Chromatography, Liquid/standards, Chromatography, Liquid/methods, Dehydroepiandrosterone Sulfate/blood, Dehydroepiandrosterone Sulfate/analysis, Dehydroepiandrosterone Sulfate/standards, Middle Aged, Liquid Chromatography-Mass Spectrometry, androstenedione, dehydroepiandrosterone sulfate, harmonization, inter-laboratory performance, liquid chromatography-tandem mass spectrometry, testosterone