Direct Cementation of Dual Mobility Cups Into the Bony Acetabulum in Primary Total Hip Arthroplasty: Clinical and Radiographic Outcomes at a Minimum 5-Year Follow-Up.
Détails
Télécharger: 36566998.pdf (573.13 [Ko])
Etat: Public
Version: Final published version
Licence: CC BY 4.0
Etat: Public
Version: Final published version
Licence: CC BY 4.0
ID Serval
serval:BIB_8A378669C2CA
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Direct Cementation of Dual Mobility Cups Into the Bony Acetabulum in Primary Total Hip Arthroplasty: Clinical and Radiographic Outcomes at a Minimum 5-Year Follow-Up.
Périodique
The Journal of arthroplasty
ISSN
1532-8406 (Electronic)
ISSN-L
0883-5403
Statut éditorial
Publié
Date de publication
06/2023
Peer-reviewed
Oui
Volume
38
Numéro
6
Pages
1120-1125
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Publication Status: ppublish
Résumé
Cemented dual mobility cups (DMCs) are commonly used in combination with acetabular reinforcement devices. Indeed, according to literature, direct cementation of metal-backed acetabular components into the bony acetabulum remains controversial as this technique is potentially associated with increased rates of aseptic loosening. Therefore, this study aimed to evaluate the clinical and radiographic outcomes of DMC cemented into the bony acetabulum in primary total hip arthroplasty (THA).
A total of 49 THA (48 patients, mean age 78 years [range, 51 to 91]) performed with direct cementation of a DMC into the bony acetabulum were prospectively included in our total joint registry and retrospectively reviewed. The clinical outcome was assessed using the Harris hip score (HHS). The radiographic outcome included measurement of component positioning and occurrence and progression of demarcation around the cemented DMC. Complications were reported with a particular attention to cemented fixation failure and aseptic loosening.
At a 7-year mean follow-up (range, 5 to 8), the pre-to postoperative HHS improved from 47 (range, 30 to 58) to 92 points (range, 80 to 98) (P < .01). Nonprogressive and focalized demarcations were observed in 7 THA (14%). Importantly, no progressive demarcation or DMC aseptic loosening was observed.
Direct cementation of DMC into the bony acetabulum ensured a stable fixation with no progressive demarcation or aseptic loosening at midterm follow-up. Therefore, this technique can be selectively considered in primary THA, especially in elderly or frail patients to avoid potential mechanical failure of press-fit fixation due to altered bone quality or additional morbidity related to the use of acetabular reinforcement devices.
A total of 49 THA (48 patients, mean age 78 years [range, 51 to 91]) performed with direct cementation of a DMC into the bony acetabulum were prospectively included in our total joint registry and retrospectively reviewed. The clinical outcome was assessed using the Harris hip score (HHS). The radiographic outcome included measurement of component positioning and occurrence and progression of demarcation around the cemented DMC. Complications were reported with a particular attention to cemented fixation failure and aseptic loosening.
At a 7-year mean follow-up (range, 5 to 8), the pre-to postoperative HHS improved from 47 (range, 30 to 58) to 92 points (range, 80 to 98) (P < .01). Nonprogressive and focalized demarcations were observed in 7 THA (14%). Importantly, no progressive demarcation or DMC aseptic loosening was observed.
Direct cementation of DMC into the bony acetabulum ensured a stable fixation with no progressive demarcation or aseptic loosening at midterm follow-up. Therefore, this technique can be selectively considered in primary THA, especially in elderly or frail patients to avoid potential mechanical failure of press-fit fixation due to altered bone quality or additional morbidity related to the use of acetabular reinforcement devices.
Mots-clé
Humans, Aged, Arthroplasty, Replacement, Hip, Acetabulum/diagnostic imaging, Acetabulum/surgery, Hip Prosthesis, Follow-Up Studies, Retrospective Studies, Cementation, Prosthesis Design, Reoperation/methods, Metals, Prosthesis Failure, aseptic loosening, cemented fixation, demarcation, dual mobility cup, total hip arthroplasty