Up to what extent doses can be lowered in pediatrict CT?

Détails

ID Serval
serval:BIB_6F00384B5C2D
Type
Actes de conférence (partie): contribution originale à la littérature scientifique, publiée à l'occasion de conférences scientifiques, dans un ouvrage de compte-rendu (proceedings), ou dans l'édition spéciale d'un journal reconnu (conference proceedings).
Sous-type
Poster: résume de manière illustrée et sur une page unique les résultats d'un projet de recherche. Les résumés de poster doivent être entrés sous "Abstract" et non "Poster".
Collection
Publications
Institution
Titre
Up to what extent doses can be lowered in pediatrict CT?
Titre de la conférence
RSNA 2007, Radiological Society of North America, 93rd Scientific Assembly and Annual Meeting
Auteur⸱e⸱s
Gutierrez D., Verdun F., Alamo Maestre L., Schnyder P., Bochud F., Gudinchet F.
Adresse
Chicago, Illinois, November 25-30, 2007
Statut éditorial
Publié
Date de publication
2007
Langue
anglais
Résumé
PURPOSE: TO determINe dose thresholds, IN term of CTDIvol, where subtle diagnostically relevant details of pediatric CT images BEcomes no more detectable.
METHOD AND MATERIALS: Abdomen images of 8 patients (age range 2 TO 16 years old) were acquired usINg a 8-slice CT system (GE Lightspeed Ultra) with CTDIvol values IN the range of 17 TO 22 mGy adapted for patient age and compatible with our national Reference Levels. Raw data were then processed TO simulate major dose reductions (UP TO tenfold). All images were reviewed and scored INdependently by four experienced radiologists usINg the VGA methodology (Visual GradINg Analysis) TO determINe the dose threshold where a signifiCANt loss of normal anaTOmy conspicuity appeared. Data were analyzed with ANOVA and Tukey HSD tests, a p>0.05 was considered TO BE signifiCANt.
RESULTS: No signifiCANt difference IN VGA scorINg appeared for CTDIvol as low as 3 mGy IN children under 10 years old (≤35 kg), whereas a signifiCANt loss of image quality was noticed when lowerINg the CTDIvol from 10 TO 3 mGy for children IN the 10 TO 16 years groUP (> 35 kg).
CONCLUSION: Compared TO the present Reference Levels proposed for pediatric CT it seems that dose CAN BE further reduced without signifiCANt loss of normal anaTOmy conspicuity. Nevertheless, a CTDIvol lower than 3 mGy IN abdomINal TOmography for children over 35 kg produces images where diagnostically relevant details are no more detectable.
CLINICAL RELEVANCE/APPLICATION: AvoidINg unnecessary dose durINg CT examINations on children while keepINg a sufficient image quality level.
Création de la notice
08/04/2008 16:47
Dernière modification de la notice
20/08/2019 15:28
Données d'usage