Enregistrement des fluctuations de la pression intraoculaire sur 24 heures par une lentille de contact intégrant un capteur: tolérance et résultats chez 10 sujets sains

Détails

Ressource 1Télécharger: BIB_6D00A408F1D1.P001.pdf (1590.12 [Ko])
Etat: Public
Version: Après imprimatur
ID Serval
serval:BIB_6D00A408F1D1
Type
Thèse: thèse de doctorat.
Collection
Publications
Institution
Titre
Enregistrement des fluctuations de la pression intraoculaire sur 24 heures par une lentille de contact intégrant un capteur: tolérance et résultats chez 10 sujets sains
Auteur⸱e⸱s
DE SMEDT Stefan
Directeur⸱rice⸱s
SCHNYDER Corinne
Détails de l'institution
Université de Lausanne, Faculté de biologie et médecine
Statut éditorial
Acceptée
Date de publication
2011
Langue
français
Nombre de pages
18
Résumé
AbstractPurpose: to evaluate the tolerability, comfort and precision of the signal transmission of an ocular Sensor used for 24-hour intraocular pressure fluctuation monitoring in humans.Patients and methods: In this uncontrolled open trial involving 10 healthy volunteers an 8.7 mm radius prototype ocular telemetry Sensor (SENSIMED Triggerfish®, Lausanne, Switzerland) and an orbital bandage containing a loop antenna were applied and connected to a portable recorder after full eye examination. Best corrected visual acuity and position, surface wetting ability and mobility of the Sensor were assessed after 5 and 30 minutes, 4, 12 and 24 hours. Subjective wearing comfort was scored and activities documented in a logbook. After Sensor removal a full eye examination was repeated and the recorded signal analyzed.Results: The comfort score was high and did not fluctuate significantly over time. The mobility of the Sensor was limited across follow-up visits and its surface wetting ability remained good. Best corrected visual acuity was significantly reduced during Sensor wear and immediately after its removal (from 1.07 before, to 0.85 after, P-value 0.008). Three subjects developed a mild, transient corneal abrasion. In all but one participant we obtained usable data of a telemetric signal recording with sufficient sensitivity to depict ocular pulsation.Conclusions: This 24-hour- trial has encouraging results on the tolerability and functionality of the ocular telemetric Sensor for intraocular pressure fluctuation monitoring. Further studies with different Sensor radii conducted on a larger study population are needed to improve comfort, precision and interpretation of the telemetric signal.