Method for confirmation of synthetic corticosteroids in doping urine samples by liquid chromatography-electrospray ionisation mass spectrometry

Details

Serval ID
serval:BIB_69D2FE367057
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Method for confirmation of synthetic corticosteroids in doping urine samples by liquid chromatography-electrospray ionisation mass spectrometry
Journal
Journal of Chromatography A
Author(s)
Fluri  K., Rivier  L., Dienes-Nagy  A., You  C. H., Maitre  A., Schweizer  C., Saugy  M., Mangin  P.
ISSN
0021-9673
Publication state
Published
Issued date
2001
Peer-reviewed
Oui
Volume
926
Number
1
Pages
87-95
Abstract
In this study, we report on the development of a method to confirm simultaneously nine of the most commonly abused synthetic corticosteroids in urine based on liquid chromatography-electrospray ionisation mass spectrometry. A considerable simplified sample preparation procedure, including liquid-liquid phase extraction with Extrelut-NT3 columns, provided both excellent sample purification and high overall recoveries. Complete HPLC separations were obtained on a reversed-phase column with 1 mM ammonium acetate-acetonitrile (60:40, v/v) as mobile phase. Mass spectral acquisition was done in the negative ion, and selected ion monitoring modes to identify the drugs with at least three characteristic ions. Detection limits were determined at less than or equal to1 ng/ml and the confirmation limits at 1 to 5 ng/ml.
Pubmed
Web of science
Create date
08/02/2008 17:44
Last modification date
20/08/2019 14:24
Usage data