Virtual versus true non-contrast dual-energy CT imaging for the diagnosis of aortic intramural hematoma.

Details

Serval ID
serval:BIB_2863FDB5AF98
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Virtual versus true non-contrast dual-energy CT imaging for the diagnosis of aortic intramural hematoma.
Journal
European radiology
Author(s)
Si-Mohamed S., Dupuis N., Tatard-Leitman V., Rotzinger D., Boccalini S., Dion M., Vlassenbroek A., Coulon P., Yagil Y., Shapira N., Douek P., Boussel L.
ISSN
1432-1084 (Electronic)
ISSN-L
0938-7994
Publication state
Published
Issued date
12/2019
Peer-reviewed
Oui
Volume
29
Number
12
Pages
6762-6771
Language
english
Notes
Publication types: Comparative Study ; Journal Article ; Multicenter Study
Publication Status: ppublish
Abstract
To assess whether virtual non-contrast (VNC) images derived from contrast dual-layer dual-energy computed tomography (DL-DECT) images could replace true non-contrast (TNC) images for aortic intramural hematoma (IMH) diagnosis in acute aortic syndrome (AAS) imaging protocols by performing quantitative as well as qualitative phantom and clinical studies.
Patients with confirmed IMH were included retrospectively in two centers. For in vitro imaging, a custom-made phantom of IMH was placed in a semi-anthropomorphic thorax phantom (QRM GmbH) and imaged on a DL-DECT at 120 kVp under various conditions of patient size, radiation exposure, and reconstruction modes. For in vivo imaging, 21 patients (70 ± 13 years) who underwent AAS imaging protocols at 120 kVp were included. In both studies, contrast-to-noise ratio (CNR) between hematoma and lumen was compared using a paired t test. Diagnostic confidence (1 = non-diagnostic, 4 = exemplary) for VNC and TNC images was rated by two radiologists and compared. Effective radiation doses for each acquisition were calculated.
In both the phantom and clinical studies, we observed that the CNRs were similar between the VNC and TNC images. Moreover, both methods allowed differentiating the hyper-attenuation within the hematoma from the blood. Finally, we obtained equivalent high diagnostic confidence with both VNC and TNC images (VNC = 3.2 ± 0.7, TNC = 3.1 ± 0.7; p = 0.3). Finally, by suppressing TNC acquisition and using VNC, the mean effective dose reduction would be 40%.
DL-DECT offers similar performances with VNC and TNC images for IMH diagnosis without compromise in diagnostic image quality.
• Dual-layer dual-energy CT enables virtual non-contrast imaging from a contrast-enhanced acquisition. • Virtual non-contrast imaging with dual-layer dual-energy CT reduces the number of acquisitions and radiation exposure in acute aortic syndrome imaging protocol. • Dual-layer dual-energy CT has the potential to become a suitable imaging tool for acute aortic syndrome.
Keywords
Aged, Algorithms, Aortic Diseases/diagnostic imaging, Female, Hematoma/diagnostic imaging, Humans, In Vitro Techniques, Male, Middle Aged, Phantoms, Imaging, Radiography, Dual-Energy Scanned Projection/methods, Reproducibility of Results, Retrospective Studies, Sensitivity and Specificity, Tomography, X-Ray Computed/methods, Acute disease, Aorta, In vitro, Tomography, x-ray computed