Criteria for establishing shielding of multi-detector computed tomography (MDCT) rooms.
10.1093/rpd/ncq100
000277738200077
20215444
Verdun
F.R.
author
Aroua
A.
author
Baechler
S.
author
Schmidt
S.
author
Trueb
P.R.
author
Bochud
F.O.
author
article
2010-04
Radiation Protection Dosimetry
1742-3406[electronic], 0144-8420[linking]
journal
139
1-3
403-409
The aim of this work is to compare two methods used for determining the proper shielding of computed tomography (CT) rooms while considering recent technological advances in CT scanners. The approaches of the German Institute for Standardisation and the US National Council on Radiation Protection and Measurements were compared and a series of radiation measurements were performed in several CT rooms at the Lausanne University Hospital. The following three-step procedure is proposed for assuring sufficient shielding of rooms hosting new CT units with spiral mode acquisition and various X-ray beam collimation widths: (1) calculate the ambient equivalent dose for a representative average weekly dose length product at the position where shielding is required; (2) from the maximum permissible weekly dose at the location of interest, calculate the transmission factor F that must be taken to ensure proper shielding and (3) convert the transmission factor into a thickness of lead shielding. A similar approach could be adopted to use when designing shielding for fluoroscopy rooms, where the basic quantity would be the dose area product instead of the load of current (milliampere-minute).
eng
60_published
true
Publication types: Journal Article ; Research Support, Non-U.S. Gov't
Publication Status: ppublish
University of Lausanne
mailto:serval_help@unil.ch
http://www.unil.ch/serval
http://serval.unil.ch/disclaimer
https://serval.unil.ch/notice/serval:BIB_2E57DF7B305B